Image in intervention

3D image fusion for live guidance of stent implantation in aortic coarctation – magnetic resonance imaging and computed tomography image overlay enhances interventional technique

Sebastian Goreczny
,
Pawel Dryzek
,
Tomasz Moszura
,
Titus Kühne
,
Felix Berger
,
Stephan Schubert

Adv Interv Cardiol 2017; 13, 3 (49): 269–272
Online publish date: 2017/09/25
Article file
- 3D image.pdf  [0.27 MB]
Get citation
 
 
Non-invasive three-dimensional (3D) imaging belongs in routine practice in diagnosis and treatment planning of many congenital and acquired cardiovascular defects. Improvements in the development of fusion imaging software have led to the introduction of 3D reconstructed images for guidance of various transcatheter therapies, with three-dimensional rotational angiography (3DRA) being the most popular tool [1, 2]. A recently available 3D roadmap based on pre-registered computed tomography (CT) or magnetic resonance imaging (MRI) data sets promises reduction in contrast and radiation exposure along with shorter procedural times [3, 4]. We present our initial experience of this application with 3D image fusion (“merging”) for live guidance of stent implantation in coarctation of the aorta (CoA), using MRI applied with the HeartNavigator prototype (Philips Healthcare, Best, The Netherlands) and CT with the commercially available VesselNavigator (Philips Healthcare).
A 12-year-old boy presented to the outpatient clinic with arterial hypertension and intermittent headache already receiving -blocker therapy. Brachiocephalic hypertension was detectable with a gradient of 47–56 mm Hg between the arms and legs, and a reduced femoral pulse quality was present. Echocardiography showed normal biventricular function and left ventricular hypertrophy. The descending aorta showed a reduced pressure profile and diastolic “run-off” with no clearly detectable gradient. An MRI scan, performed with acquisition of a 3D whole heart sequence, demonstrated subatretic CoA distal to the left subclavian artery (LSA) with increased collateral flow. The aortic arch measured 10 × 12.5 mm, CoA diameter was 5 × 5 mm and the descending aorta distal to the narrowing was 17 × 19 mm.
Interventional therapy was performed under conscious sedation via femoral artery access. The three-dimensional whole heart sequence was uploaded to the workstation (HeartNavigator prototype, Philips), automatically segmented and manually corrected (Figure 1 A). For accurate fusion with live fluoroscopy the roadmap was manually aligned with two angiographies (10–15 ml) performed at a minimum 30° difference in angulation from the anterior-posterior projection (Figure 1 B). Initial hemodynamic measurements confirmed significant stenosis with a systolic pressure gradient of 50 mm Hg. A 28 mm long covered Cheatham Platinum (CP) stent (NuMed, USA) was mounted on a 12 mm Balloon-in-Balloon (BIB) catheter (NuMed), delivered through an 11 Fr long vascular sheath (Cook, USA) and positioned without additional contrast injections (Figures 1 C, D). After postdilation with an 18 mm sizing balloon (SJM/AGA Medical, USA), a 10 mm Hg residual gradient was left for re-dilatation 6 months later. The fluoroscopy time was 4.3 min and the dose area product was 15 940 mGray • cm².
A 7-year-old boy was referred for treatment for recently diagnosed CoA. Brachiocephalic hypertension was detectable with a gradient of 37–42 mm Hg between the arms and legs, and a reduced femoral pulse quality was present. Echocardiography showed normal left and right ventricular function with marked left ventricular hypertrophy. Distal to the origin of the LSA, flow acceleration was noted with peak gradient of 38 mm Hg (Vmax = 2.8 m/s). The descending aorta showed a reduced pressure profile and diastolic “run-off”. A chest contrast CT scan confirmed the diagnosis of critical CoA distal to the LSA with well-developed collateral vessels. The aortic arch and the stenosis measured 11 × 12.5 mm and 2 × 2 mm, respectively.
The raw CT dataset was uploaded to the dedicated workstation (Philips Workspot), manually segmented and prepared for fusion before the patient’s arrival at the catheterization laboratory (Figure 2 A). The target lesion was re-measured and ring markers were placed to indicate the ostia and define the landing zone for stent implantation.
The treatment was performed under general anaesthesia via femoral artery access. Short fluoroscopy sequences in anterior-posterior and left lateral projections were used for 2D–3D fusion (Figure 2 B). Vertebral bodies (white stars) of the mid and lower thoracic spine served as reference points. Initial hemodynamic measurements showed a systolic gradient of 35 mm Hg across the stenosis. For safer introduction of a large, long vascular sheath (12 Fr, Cook) the lesion was predilated with a high pressure balloon (Figure 2 C). Next, a 34 mm covered CP stent (NuMed) was mounted on a 16 mm BIB catheter (NuMed), delivered and implanted solely under 3D guidance (Figure 2 D). The blue marking ring placed at the origin of the LSA was used as a marker for the distal end of the stent. Final rotational angiography confirmed perfect stent opposition to the walls of the aorta, without any leak and with unobstructed flow to the LSA. Hemodynamic measurements showed no residual gradient. The fluoroscopy time was 4.2 min, and the dose area product was 12269 mGray • cm².
In the majority of patients with native aortic coarctation and in almost all with recoarctation, trans-catheter intervention is the first line of treatment. When guided with traditional 2D angiography, it is advised to acquire at least two perpendicular projections to visualize the lesion and perform pertinent measurements. In some patients, standard anterior-posterior and lateral projections may not be sufficient to completely profile the stenosis, or to precisely identify the origin of head and neck vessels. With recent improvements in imaging software, 3D mapping can be performed with CT or MRI data implementation [3, 4]. Three-dimensional rotational angiography provides a large number of “dynamic” angiographic and “static” reconstructed images for the complete evaluation of the lesion; however, it requires specific setup of the angiographic system, injection of a relatively large volume of contrast and additional radiation exposure [1, 2]. These disadvantages may be avoided with a 3D roadmap and implementation of CT or MRI data.
With recently available 3D image fusion software, both CT and MRI datasets may be used for easy segmentation and production of a 3D reconstruction of the target lesion prior to the patient’s arrival at the catheterization laboratory [3, 4]. The option of 2D–3D registration eliminates the need to perform a rotational spin and therefore simplifies and shortens the registration process. Reliable overlay omits the need for diagnostic angiography and enhances precise stent positioning and implantation. Three-dimensional mapping might be supported, if it increases overall efficacy and safety of complex cardiovascular interventions. This, and the potential reduction in contrast and radiation exposure, has to be proven in larger series.
Three-dimensional mapping of congenital and structural heart disease may improve the interventional quality and process with the implementation of CT or MRI datasets to fluoroscopy. This initial experience provides proof of concept for a multi-center approach and supports the need for a prospective study. Mapping from CT and MRI can be defined as beneficial in comparison to 3DRA, which is still the standard for 3D roadmap application.

Acknowledgments

The authors acknowledge Dr Maciej Lukaszewski and Ms Alexandra Krawczuk from the Polish Mother’s Memorial Hospital, for their contribution to this study.

Conflict of interest

The authors declare no conflict of interest.

References

1. Goreczny S, Dryzek P, Moszura T, et al. Rotational angiography in monitoring of covered CP stent implantation in patient with critical aortic coarctation and patent ductus arteriosus. Kardiol Pol 2012; 70: 505-7.
2. Glöckler M, Halbfa J, Koch A, et al. Multimodality 3D-roadmap for cardiovascular interventions in congenital heart disease: a single-center, retrospective analysis of 78 cases. Catheter Cardiovasc Interv 2013; 82: 436-42.
3. Goreczny S, Dryzek P, Moszura T. Use of pre-intervention imaging with a novel image fusion software for guidance of cardiac catheterisation in a patient with pulmonary atresia and major aortopulmonary collaterals. Cardiol Young 2016; 26: 1438-40.
4. Goreczny S, Dryzek P, Moszura T. Novel 3-dimensional image fusion software for live guidance of percutaneous pulmonary valve implantation. Circ Cardiovasc Interv 2016; 9: e003711.
Copyright: © 2017 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.