Usefulness of the Evolution mechanical dilator sheath for endocardial lead extraction – preliminary results

Andrzej Przybylski
,
Paweł Syska
,
Artur Oręziak
,
Aleksander Maciąg
,
Maciej Steriński
,
Krzysztof Kuśmierski
,
Mariusz Kuśmierczyk

Post Kardiol Interw 2010; 6, 2 (20): 59-65
Online publish date: 2010/06/28
Article file
- Usefulness.pdf  [0.34 MB]
Get citation
 
 

Introduction
There has been a growing number of patients requiring endocardial lead extraction in recent years. This is related to:
• increased number of implanted pacemakers, cardioverer-defibrillators (ICD) and resynchronization systems (cardiac resynchronization therapy, CRT) [1],
• increased life expectancy in patients with implanted devices, which causes an increased number of device replacement or upgrading procedures (to ICD or CRT),
• igrowing number of cases of damaged leads (especially defibrillating leads).
Despite development of new techniques of lead extraction the procedure is still related to high risk of serious complications or procedure failure [2-5]. This is mainly caused by the presence of adhesions between the lead and the cardiovascular system. Different methods including mechanical force, radiofrequency energy or laser techniques have been used to separate leads from the surrounding connective tissue [2, 6-8]. Mechanical methods are still the most commonly used. This is caused by a longlasting experience in their application, high availability of the devices and economic aspects. Moreover, some studies reported similar results for these devices in comparison to newer techniques [9-11].
One of the newest mechanical methods is the use of Evolution sets (Cook Medical Inc. Bloomington, USA). Their inventiveness comes from the fact that the inner sheath is ended with a metal ring, which can be given a rotational force by the use of a handle resembling a gun trigger (fig. 1). Rotational movement of the sheath ended with a metal ring (fig. 2) is used to liberate the lead from the fibrous tissue. There have been no Polish reports on the use of the Evolution system. Besides, only single reports on its application can be found in the literature [12]. The aim of the study was to present our own, first in Poland experiences with the lead extraction Evolution system.
Materials and methods
Studied group

The analysis included a group of consecutive 45 patients who underwent a percutaneous extraction of an endocardial lead between January 2008 and April 2010 (during that period the Evolution system was available at our site). Qualification for the procedure was based on the current guidelines established by cardiologic societies [2]. Patients baseline characteristics are presented in table 1. Indications for the procedure included mainly infective complications related to implanted devices, but also lead damage and in some cases the need of system upgrade – table 2.
Procedure description
Procedures were set in the cardiosurgical operating room with possibility of emergency sternotomy and surgery with the use of extracorporeal circulation. All patients signed an informed consent for the procedure and for general anaesthesia. Intratracheal general anaesthesia was used. ECG, blood and central venous pressure and pulsoxymetry were monitored continuosuly throughout the procedure. All steps of the procedure related to manouvers inside the cardiovascular system were performed under direct fluoroscopy. Patients in need of permanent pacing had a temporary lead introduced through the femoral or cubital fossa vein contralateral to the side of the previously implanted device.
After opening of the pacemaker/ICD pocket and removal of the pulse generator, leads were mobilised from ligatures with the use of an electric knife. In case of removal of more the one lead, left ventricular lead was extracted first (from the cardiac veins system), followed by removal of an atrial and a right ventricular lead. After liberation of a lead from ligatures and connective tissue adhesions inside the pocket, a standard leader was introduced through the lumen of a lead to assess the patency of the central canal necessery for the placement of the locking-stylet leader (for example Liberator, Cook). To disconnect the actively fixed lead from the endocardium the fixation element was unscrewed under direct fluoroscopy. These manouvers were followed by an attampt to extract the lead with the use of manual traction. If manual traction had failed leads were removed percutaneously with the use of a set of metal, polypropylene and teflon Byrd dilators (Cook). After adequate preparation of the lead (cutting of the IS-1 tip, dilation of the central canal ostium, placement of the locking-stylet leader and long ligatures at the isolation of the distal tip of a lead which were than used to give tension on the lead during introduction of a dilator) an attempt was made to disconnect the lead from the connective tissue adhesions throughout its way in the venous system with the use of adequate dilators. Steinless steel sheaths were used to reduce the resistance between the clavicula and the first rib. Its use seemed especially needed in case of calcifications of the costoclavicular ligament. After elimination of resistance related to strangulation of a lead in the proximity of the costo­clavicular ligament, the steinless steel sheaths were retracted and the new set of telescope sheaths adjusted to the size of the lead (7-16 F) was introduced. During introduction of successive sheaths lead was mildly pulled with the use of locking-stylet leader and ligatures placed around its corpus. A decision to use Evolution system (Cook) was taken in case of unsuccessful separation of the lead from adhesions with the use of teflon or poly­propylene sheaths, especially on its passage through the subclavian and innominate vein.
If a free tip of an lead was situated inside the vascular system preventing its capture via an upper access, a lower access via a femoral vein was used by introduction of a teflon sheath ended with a special loop enabling capture of the lead tip and its introduction inside the sheath (Needles-Eye, Cook). After capturing of the lead and its placement inside the sheath its was possible to perform a counter-traction to liberate the lead tip from the endocardium. Success of the procedure and procedure complications were defined according to current recommendations of the cardiologic societies.
Statistical analysis
Continuous variables are presented as arithmetic means ± standard deviations (SD) and categorical variables as numbers and percentages (%).
Results
In the studied group of 45 patients extraction of 76 leads was attempted. Data on the type and number of removed leads are presented in table 3.
Extracted leads consisted mainly of pacemaker leads, but in 8 cases defibrillating lead were removed. The proportion of actively or passively fixed leads was similar with insignificant excess of actively fixed leads.
Two leads were extracted at the same time in half of the patients. Three patients underwent extraction of 3 leads. Fully successful procedure defined as a complete lead extraction was obtained in 95% of cases.
Techniques used
Direct traction allowed extraction of 31 (41%) leads. In the remaining cases the counter-traction method with application of the Cook system elements was used (tab. 4). The oldest leads extracted with the use of a direct traction had been implanted 5 years before. These were pacing leads and one of them was a passive fixation lead.
The use of the Evolution system
The Evolution system was used for extraction of 6 leads (8% – 2 defibrillating leads and 4 pacing leads) in 3 pa­tients. This technique was applied only after separation of leads from adhesions had been impossible with the use of Byrd sheaths. Five leads were completely extracted. One defibrillating lead was not completely removed, because its adherence to the innominate vein could not have been overcome. This situation occurred in a 22-year-old female patient after an ICD implantation 4 years before in the secondary profilaxis of sudden cardiac death.
Complications
Complications were classified according to definitions proposed by cardiologic societies [2]. There were no serious complications observed in the analyzed group (death, the need for surgical intervention, pulmonary embolism, complications related to anaesthesia, stroke, infection of the device implanted on the contralateral side).
One patient required blood transfusion due to procedural blood loss. This patient was subjected to extraction of two pacing leads implanted 3 years before with subsequent implantation of CRT-D. Telescope sheaths and Evolution system were used for lead removal in this case.
Discussion
We present for the first time in Poland the application of a new method of endocavitary lead extraction based on the Evolution system. This system was used for extraction of 6 out of 76 leads (8%) in the analyzed group. In all cases the technique was applied after extraction with means of other methods had failed. Five leads were completely extracted and there was one extraction failure. There was no serious complications related to the use of the Evolution technique.
This method was a subject of only few reports found in the current literature [12]. It may be partially related to a short time which has passed since the introduction of this system. The largest group of procedures performed with the use of the Evolution system was presented by Hussein et al. [12]. This technique was used in 29 patients, in whom extraction of 41 leads was attempted in years 2008-2009. The Evolution system was used as a method of choice for lead extraction in 12 patients (16 leads). In this group of patients 100% success rate was obtained. In the remaining cases (17 patients, 25 leads) the Evolution system was used after other techniques had prooven unsuccesfull. In this group of patients the rate of successful extractions was lower (77%). Application of additional techniques was necessary in 4 patients (the use of a loop introduced via a femoral vein in 2, laser sheaths in 2).
Our results are similar to the findings presented in the cited report. We used the Evolution system solely in cases of unsuccessful lead extraction with means of other techniques (direct traction, Byrd sheaths), which happens frequently when the structure of a lead is damaged making the procedure more difficult. Rare use of the Evolution system is mainly related to a relatively high price of the system. It should be mentioned that lead extraction procedures performed for indications other than infective endocarditis are not reimbursed by the National Health Fund.
Of note, infective complications (infection of the pocket, infective endocarditis) were an indication for the procedure in only 42% of patients. In the remaining cases indication for the extraction consisted of lead damage or the need for removal of unnecessary leads prior to system upgrade (for example an upgrade of pacing device to CRT-D). Is seems likely that these indications will predominate in the near future [6, 9, 13]. It is related to the extension of indications for multi-lead device implantation (such as CRT-D) and frequent lead damage (especially defibrillating leads). Another problem is venous obstruction at the side of implantation which prevents further lead implantations. Venous obstruction is caused mainly by the adhesions between the lead and the vascular wall. Their presence is a cause of the most severe difficulties related to lead extraction [10, 14]. Adhesions are often found in the subclavian or/and innominate vein, particularly in young people. Liberation of a lead is also difficult when proximal defibrillating coil is situated in the subclavian vein. The fact that it was impossible to remove the proximal part of the lead in 4 patients during heart transplantation in our group best exemplifies how strongly can the lead adhere to the vascular wall. These patients required extraction of remaining lead fragments with the use of described methods. In two of them extracted lead was a defibrillating dual-coil type and in the third case a pacing lead.
On the other hand, in some patients it was possible to extract leads with means of direct traction even long time after implantation (41% of leads). These situations included leads implanted 5 years before. Similar results were presented by other authors. Using this technique Kutarski et al. extracted 33 leads implanted at least one year before [9]. Therefore it seems resonable to attempt a direct traction method in each case of lead extraction [7, 15]. It should be remembered that direct traction must be gentle in order not to damage the lead structure, which makes the subsequent removal difficult.
Study limitations
It was impossible to compare the use of the Evolution system with other, newer techniques of lead extraction (laser, radiofrequency current), because they were not available.
In our study the Evolution system was applied only after failed extraction by means of other methods and therefore we were unable to compared times of procedure or fluosroscopy.
Conclusion
The mechanical Evolution system is a safe method of lead extraction and increases the effectiveness of the procedure.

References
1. The EHRA White Book. The Current Status of Cardiac Electrophysiology in ESC Members Countries. http://www.escardio.org/communities /EHRA.  
2. Wilkoff BL, Love CJ, Byrd CL, et al. Heart Rhythm Society; American Heart Association. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm 2009; 6: 1085-1104.  
3. Smith MC, Love CJ. Extraction of transvenous pacing and ICD leads. Pacing Clin Electrophysiol 2008; 31: 736-752.  
4. Wilkoff BL, Byrd CL, Love CJ, et al. Pacemaker lead extraction with the laser sheath: results of the pacing lead extraction with the excimer sheath (PLEXES) trial. J Am Coll Cardiol 1999; 33: 1671-1676.  
5. Kennergren C, Bucknall CA, Butter C, et al. PLESSE investigators group. Laser-assisted lead extraction: the European experience. Europace 2007; 9: 651-656.  6. Kennergren CA. European perspective on lead extraction: part I. Heart Rhythm 2008; 5: 160-162.  
7. Kennergren CA. European perspective on lead extraction: part II. Heart Rhythm 2008; 5: 320-323.  
8. Neuzil P, Taborsky M, Rezek Z, et al. Pacemaker and ICD lead extraction with electrosurgical dissection sheaths and standard transvenous extraction systems: results of a randomized trial. Europace 2007; 9: 98-104.  
9. Kutarski A, Małecka B, Ruciński P, et al. Percutaneous extraction of endocardial leads – a single centre experience in 120 patients. Kardiol Pol 2009; 67: 149-156.
10. Bongiorni MG, Soldati E, Zucchelli G, et al. Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads. Eur Heart J 2008; 29: 2886-2893.
11. Hussein AA, Wilkoff BL, Martin DO, et al. Initial experience with the Evolution mechanical dilator sheath for leadextraction: safety and efficacy. Heart Rhythm 2010 Mar 23.
12. Calvagna GM, Evola R, Scardace G, et al. Single-operator experience with a mechanical approach for removal of pacing and implantable defibrillator leads. Europace 2009; 11: 1505-1509.
13. Jones SO 4th, Eckart RE, Albert CM, et al. Large, single-center, single-operator experience with transvenous lead extraction: outcomes and changing indications. Heart Rhythm 2008; 5: 520-525.
14. Novak M, Dvorak P, Kamaryt P, et al. Autopsy and clinical context in deceased patients with implanted pacemakers and defibrillators: intracardiac findings near their leads and electrodes. Europace 2009; 11: 1510-1516.
15. Kennergren C, Bjurman C, Wiklund R, Gäbel J. A single-centre experience of over one thousand lead extractions. Europace 2009; 11: 612-617.
Copyright: © 2010 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.