Introduction
In recent years, the number of catheter-based endovascular procedures has gradually increased. In particular, novel transarterial locoregional anticancer treatments are increasingly being used, including gastrointestinal bleeding embolization, bland embolization, chemoembolization, and radioembolization (RE) [1].
A large variety of guidewire-directed microcatheters are commercially available today for such embolizations. The choice of microcatheter for catheterization of a particular target vessel depends on distal trackability, support, and pushability but also on the interventionalist’s preference. In theory, all microcatheters can be shaped or reshaped with steam. In certain challenging endovascular procedures, steam shaping of the catheter tip may facilitate target vessel catheterization or provide the support needed to achieve a stable tip position for embolization [2]. However, steam shaping is time-consuming and therefore rarely done in clinical practice. Moreover, the steamed catheter end often recovers its initial shape [3].
A number of microcatheters with preshaped tips are available that improve selective catheterization of arteries originating at acute angles and help maintaining position within an aneurysm during coil embolization [4, 5]. The range of shapes includes 45° or 90° angled tips, and J-, S- or C-shaped curves. The choice of a particular shape is related to the local anatomy and the intended method of treatment. However, these preshaped microcatheters are devoted mainly to neuroradiological interventions, which are performed in tortuous arteries and require a firm support during the treatment. On the other hand, oncological interventions in the visceral arteries require less support but more flexibility and steerability. The Direxion (Boston Scientific Corp., Natick, USA) is a new steerable microcatheter dedicated to visceral radiological interventions. It comes with various tip shapes and its design includes a proprietary Nitinol hypotube for steerability. Theoretically, guidewireless technique may reduce the time and cost of the procedure. In a recent comparative preclinical study on a porcine model, vessel catheterization without a guidewire was significantly more successful with the Direxion than with the Renegade (Boston Scientific, Natick, USA) steam-shaped catheter [6].
Aim
We conducted a clinical trial to verify the safety and efficacy of this new guidewireless catheterization technique in 50 consecutive patients scheduled for routine clinical yttrium-90 RE.
Material and methods
Fifty consecutive patients (32 male, 18 female; age 63.7 ±10.7 years) scheduled for RE were prospectively enrolled in this multicenter clinical trial including three centers and three operators. The study was approved by the institutional review boards, and written informed consent was obtained from each patient. All patients underwent successful angiography with selective visceral catheterization, protective embolization of extrahepatic vessels originating from liver arteries to avoid migration of microspheres to the gastrointestinal tract, and treatment simulation with Tc-99m macroaggregated albumin (MAA). For hepatic yttrium-90 delivery, the femoral artery was cannulated under local anesthesia followed by placement of a 5 Fr sheath (Introducer II; Terumo Europe, Leuven, Belgium). The celiac trunk was catheterized, depending on the degree of its outlet angle, using a 5F Cobra (Imager II C2; Boston Scientific, Natick, MA, USA), a Soft-VU (Sos Omni Selective; Angiodynamics, Queensbury, NY, USA), or a Sidewinder II (Cordis, Miami, FL, USA) catheter. Liver target vessels were catheterized coaxially with a 0.021” Direxion (Boston Scientific Corp., Natick, USA) with a preshaped J-tip. This was attempted using guidewireless technique just by utilizing the ability to steer the tip of the microcatheter. The microcatheter was advanced during injection of contrast medium to optimize fluoroscopic guidance and to prevent trauma to target vessels. In cases when the optimal microcatheter position could not be reached using guidewireless technique, a microguidewire was used.
The interventional procedure was documented recording the following information: anatomy of the liver vessels according to the classification of Michels [7], type and anatomical position of the guiding catheter, anticipated catheter position for yttrium-90 delivery, success of the catheterization technique, degree of difficulty in maneuvering the microcatheter to the target position, time to reach the target position, and any complications. Each procedure was described by the interventionalist using a standardized questionnaire (Appendix 1) designed by our team.
Results
The procedural features of the 57 RE treatments in the 50 study patients are summarized in Table I. The vast majority of patients (82%) had standard vascular anatomy (Michels I). Cobra II and Soft VU were the most commonly chosen guiding catheters, and they were typically placed in the celiac trunk.
The target position was reached in 53 RE treatments (93%), and the mean time to reach the target position was 53.3 ±53.0 s (range: 9–267 s). The reasons for unsuccessful catheterization with Direxion in four procedures were extremely tortuous vessel anatomy (n = 2; Photo 1), very large parent artery diameter (7.1 mm, n = 1), which impeded precise steering because of the lack of support from the vessel wall (Photo 2), and celiac trunk occlusion (n = 1; Photo 3).
Yttrium-90 microspheres were delivered into the right hepatic artery in 53% of cases, the left hepatic artery in 33%, the middle hepatic artery in 7%, and a segmental artery in 7%. The technical difficulty to reach the target position was rated as easy with a mean score of 2.3 ±1.3 (easy). The only complication associated with use of the Direxion catheter was a mild vasospasm in one instance.
The failure rate was too small for a statistical analysis of predisposing factors. In Spearman’s rank correlation test, successful outcome was related only to the subjective difficulty score, which is an obvious conclusion. In the Kruskal-Wallis test, difficulty score, guiding catheter selection, and target vessel were statistically related to catheterization time (p < 0.0001, p < 0.001, and p < 0.01, respectively; Figures 1 and 2). Multiple regression analysis showed that guiding catheter selection and difficulty score were independent predictors of microcatheterization time. On the other hand, in receiver-operating characteristic (ROC) curve analysis, only target vessel was a significant predictor of the catheterization time but the area under the curve (AUC) value was only of borderline significance (AUC = 0.639, 95% CI: 0.501–0.762).
Discussion
This study was the first to clinically document that the use of the Direxion microcatheter for guidewireless radioembolization is safe and has a very high technical success rate.
Liver-directed therapies for unresectable hepatic cancer are still evolving [8, 9]. These treatments involve complex endovascular procedures that generally require the use of microcatheters. Particularly in RE, successful catheterization with prophylactic embolization of small and angled vessels, such as the right gastric artery, is crucial to prevent subsequent radiation-induced gastritis [10]. In contrast, therapeutic administration of yttrium-90 microspheres through the liver arteries is technically a rather simple interventional procedure. However, it still requires the use of a microcatheter for controlled delivery of the radionuclide.
Guidewireless catheterization is still uncommon in clinical practice, as it is associated with a significant risk of vessel injury and also of kinking of the catheter in its unsupported portions. Therefore, wireless microcatheterization has been considered an off-label procedure so far. However, in a preclinical study on a porcine model Dudeck found that catheterization without a guidewire with the Direxion was significantly more successful (88%) compared with the Renegade (32%) and also significantly faster than guidewire-assisted vessel catheterization with the Renegade (16.1 vs. 27.1 s on average) [6]. Another wireless microcatheter is SwiftNINJA (Merit Medical, South Jordan, UT, USA), which has so far been successfully used in single cases of kidney angiomyolipoma embolization with ethanol and iodized oil, in arterial infusion chemotherapy with coil embolization of maxillary cancer [11], and in a small multicenter trial of various types of procedures [12]. A unique feature of SwiftNINJA is its direct ergonomic steering dial that allows the catheter tip to be deflected from straight to 180° in opposing directions. However, this device requires further clinical investigation before wider use. The third steerable microcatheter which is available on the market is Enzo (Micrus Endovascular Corp, San Jose, California, USA). It was designed for intracranial applications and has a tip that can be deflected up to 90° in opposing directions. However, there is only one published case report on the use of this microcatheter [13].
There appears to be a wide range of potential applications for steerable microcatheters in interventional radiology. Since Direxion microcatheters are currently only available for 0.021” and 0.027” systems, their application in neurovascular procedures may be limited to the delivery of stents and liquid embolics. Peripheral applications of the Direxion microcatheter seem to be much wider as it clearly enhances vessel selectivity in challenging anatomies where vessels cannot be accessed with standard systems [6]. Another potential advantage of Direxion is better control of coil loops during embolization through more precise microcatheter tip manipulations. Thus, positioning and packing of coils may be improved, reducing the risk of coil migration. Moreover, in the case of catheter kick-out during coil placement, repositioning of the microcatheter without a guidewire should be easier. The Direxion technology is also very promising for peripheral liquid embolization with glue or Onyx. A common problem in such procedures is that significant repositioning of a microcatheter filled with an embolic agent is almost impossible since a guidewire cannot be inserted at this stage. In such situations, the use of a steerable microcatheter would be invaluable. The potential to reposition the microcatheter wirelessly could also be useful in patients undergoing RE, where radionuclide during workup angiography and the radioembolic agent for treatment need to be administered from different catheter positions. In such cases, the radiologist may be able to control the application and thus eliminate the risk of radioactive contamination that may otherwise result from insertion and removal of the guidewire.
The study has some limitations. Firstly, we treated a small random sample of patients from a general population with unselected anatomy. It is obvious that anatomy is a crucial factor influencing the success rate of target vessel catheterization, as seen in two of the unsuccessful cases in our study. However, the distribution of anatomic variants in our study population (Table I) is quite close to that of the general population. Secondly, our study population was also unselected regarding target vessels. Further studies should investigate and compare outcome of RE with the Direxion microcatheter in “easy” (right and left hepatic artery) and “demanding” (segmental arteries and the right gastric artery) vascular territories. Finally, in our study, we assessed only the immediate technical success rate without follow-up to detect potential later complications of the wireless technique. Nevertheless, our experience indicates that potential arterial injury that may be a complication of the use of a wireless catheter tends to be an immediate complication, as was the vasospasm seen in one of our patients.