eISSN: 1897-4295
ISSN: 1734-9338
Advances in Interventional Cardiology/Postępy w Kardiologii Interwencyjnej
Current issue Archive Manuscripts accepted About the journal Editorial board Abstracting and indexing Subscription Contact Instructions for authors Publication charge Ethical standards and procedures
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank
3/2018
vol. 14
 
Share:
Share:
Image in intervention

Lead externalization – a potential source of lead-related infective endocarditis?

Maciej Polewczyk
,
Dorota Nowosielecka
,
Konrad Tomków
,
Łukasz Tułecki
,
Andrzej Kutarski
,
Anna Polewczyk

Adv Interv Cardiol 2018; 14, 3 (53): 316–317
Online publish date: 2018/09/21
Article file
- lead.pdf  [0.10 MB]
Get citation
 
 
A 32-year-old patient with a single chamber implantable cardioverter-defibrillator (ICD) was admitted to a cardiological ward due to recurrent episodes of fever since 1 month before. Blood tests showed a slightly elevated C-reactive protein (CRP) level without leukocytosis or procalcitonin increase. The transthoracic echocardiography (TTE) examination revealed a vegetation (29 mm × 6 mm) in the lower right atrium connected with the lead; the image was confirmed in transesophageal echocardiography (TEE). Negative blood culture was only probably negative due to the antibiotic therapy. The patient was diagnosed with lead-related infective endocarditis (LRIE), received modified antibiotic therapy and was initially qualified for the surgical extraction of the ICD system, but he did not agree to surgery. After consultation with the referential transvenous lead extraction (TLE) center the patient was qualified for TLE. Preoperative fluoroscopy revealed an interesting finding – ICD lead externalization was observed at the same location as the previously described vegetation. The TLE procedure was performed under general anesthesia in a hybrid operating room with intraoperative TEE monitoring and cardiac surgical standby. Due to the presence of a relatively large vegetation, a special nitinol basket was introduced into the main pulmonary artery for pulmonary circulation protection. The lead was removed using a Byrd dilator sheath (Cook, USA). Although during the procedure the vegetation was fragmented (only small parts were caught by the basket) there were no clinical signs of pulmonary embolism. There were no periprocedural complications. Postoperative examination of the extracted Riata (St. Jude Medical, USA) defibrillating lead showed significant cable externalization with remnants of the vegetation and relevant lead degeneration (Figure 1).
Lead externalization is relatively common among Riata leads. However, in the literature we may find cases of such a phenomenon in different leads produced by different manufacturers [1]. The mechanism of externalization of Riata leads is related to their specific construction. Inside there are several conductors which may move relative to the insulation, leading to its further damage – ‘inside-out abrasion’. In some research there was no evident correlation between lead externalization and electrical failure [2], but a large Canadian registry confirmed the impact of externalization on lead failure [3]; in our case the electrical parameters of the ICD system remained stable. The influence of lead externalization on LRIE development has not been investigated yet; there are only a few similar reports [4]. An externalized cable might be thrombogenic. A thrombus formation around the location of externalization may become the origin of a further infection. According to the recent reports, some patients with vegetations even exceeding 4 cm in diameter underwent TLE, avoiding cardiac surgery. However, the decision of the extraction method was made individually considering the total vegetation volume, the degree of infiltration of heart structures and the real risk of severe pulmonary embolism [5]. In clinical practice the decision of removal of a functional lead with present externalization remains problematic. Apart from possible electrical failure one should have in mind the fact that exposed internal lead conductors might be the chief sites of serious infection.

Conflict of interest

The authors declare no conflict of interest.

References

1. De Maria E, Borghi A, Bonetti L et al. Externalized conductors and insulation failure in Biotronik defibrillator leads: history repeating or a false alarm? World J Clin Cases 2017; 5: 27-34.
2. Demirel F, Adiyaman A, Delnoy PP, et al. Mechanical and electrical dysfunction of Riata implantable cardioverter-defibrillator leads. Europace 2014; 16: 1787-94.
3. Parkash R, Thibault B, Mangat I, et al. Canadian Registry of implantable electronic device outcomes: surveillance of the Riata lead under advisory. Circ Arrhythm Electrophysiol 2016; 9: pii: e004282.
4. Palamà Z, Trotta R, Mandurino C, et al. Could externalized St. Jude Medical Riata® lead be a culture medium of a polymicrobial endocarditis? A clinical case. Case Rep Cardiol 2017; 2017: 8967234.
5. Polewczyk A, Jachec W, Tomaszewski A, et al. Lead-related infective endocarditis: factors influencing the formation of large vegetations. Europace 2017; 19: 1022-30.
Copyright: © 2018 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.