Bieżący numer
Archiwum
Filmy
Artykuły w druku
O czasopiśmie
Suplementy
Rada naukowa
Recenzenci
Bazy indeksacyjne
Prenumerata
Kontakt
Zasady publikacji prac
Opłaty publikacyjne
Standardy etyczne i procedury
Panel Redakcyjny
Zgłaszanie i recenzowanie prac online
|
3/2021
vol. 123 streszczenie artykułu:
Artykuł oryginalny
The accuracy of intraocular lens power calculation formulas for eyes longer than 25.00 mm and the correlation between method accuracy and eyeball axial length
Wiktor Stopyra
1
KLINIKA OCZNA 2021, 123, 3: 129–133
Data publikacji online: 2021/10/13
Pełna treść artykułu
Pobierz cytowanie
ENW EndNote
BIB JabRef, Mendeley
RIS Papers, Reference Manager, RefWorks, Zotero
AMA
APA
Chicago
Harvard
MLA
Vancouver
Aim of the study To compare the accuracy of intraocular lens power calculation formulas and examine the correlation of this accuracy with the axial length for eyes larger than 25.00 mm Material and methods The data of myopic patients who underwent uneventful phacoemulsification between October 2015 and June 2019 were reviewed. The intraocular lens power for each patient was calculated using 6 formulas (Holladay1, SRK/T, Hoffer Q, Holladay 2, Haigis and Barrett Universal II) before cataract surgery. Postoperative refraction was measured and refractive prediction error was calculated 3 months after phacoemulsification. The correlation between axial length and absolute error was evaluated. Results Seventy patients (81 eyeballs with ocular axial length between 25.01 mm and 28.57 mm) were studied. The Barrett Universal II formula achieved the lowest level of mean absolute error of 0.08 ±0.08 D (p < 0.001). Significant correlations for the Holladay 1 (r = 0.478, p < 0.001), the Barrett Universal II (r = 0.312, p = 0.005) and for the Hoffer Q (r = 0.293, p = 0.008) formula were observed. Conclusions 1. The Barrett Universal II formula obtained the lowest absolute error and is recommended for intraocular lens power calculation for eyeballs with axial length exceeding 25.0 mm. 2. The correlation between axial length and absolute error is a factor which should be considered when calculating intraocular lens power, especially for eyeballs with an extremely large axial length. |
|