Calcified plaques are associated with procedural challenges and suboptimal outcomes of percutaneous coronay intervention (PCI), which can result in impairment of stent apposition and stent underexpansion. Good results were observed after laser atherectomy, rotational atherectomy or high-pressure balloon dilation [1]. Intravascular lithotripsy (IVL) is a novel alternative in plaque modification. It is effective, safe and easy to perform in severely calcified segments [2]. The coronary IVL catheter is a single-use device that contains lithotripsy emitters enclosed in an integrated balloon. The emitters generate sonic pressure waves creating a field effect to treat vascular calcification. The IVL catheter is available in 2.5 to 4.0 mm diameters with 12 mm in length and delivers 10 pulses/10 s with a maximum of 80 pulses on one balloon catheter [3].
We present images from angiography and optical coherence tomography of a 58-year-old man who was admitted to the hospital in a tertiary center with anterior-wall ST-segment elevation myocardial infarction. He complained of chest pain from 30 min, but felt angina symptoms in the last days after walking 100–150 m. The coronary angiogram revealed severely calcified critical stenosis of the proximal and middle left anterior descending artery (LAD). Several inflations with non-compliant (NC) balloons (2.5–3.0 mm) were done and two 2.75 mm drug-eluting stents (DES) were implanted in the LAD. Unfortunately the proximal stent was unexpanded. A decision was made to transfer the patient to our clinic to perform IVL with a Shockwave device (Shockwave Medical, Fremont, California, United States).
We performed angiography and OCT to evaluate stent expansion and assess the size of the plaque calcification (Figures 1 A, B). Next we decided to use the IVL Shockwave C2 catheter (3.0 × 12 mm). Four sessions with ten applications were performed. After the first pulses the balloon remained partially unexpanded in the proximal stent, but further cycles were done and the NC balloon (3.25 × 12 mm) allowed us to achieve full dilation of the implanted stent (Figure 1 C). Optical coherence tomography (OCT) showed complete stent apposition with dissection in the ostial LAD (Supplementary Figure S1). It was treated with one DES (3.5 × 15 mm) and post-dilatation with an NC balloon (3.5 × 12 mm) after stent deployment was done. Control angiography and OCT confirmed a good PCI outcome with optimal stent expansion and apposition (Figure 1 D). The patient was discharged 2 days later without complications. A mean lumen diameter of proximal edge of stent was 1,66 mm and a minimal lumen area of 2,21 mm2. After intervention with intravascular lithotripsy OCT confirmed optimal stent expansion and apposition, with a minimal lumen area 9,3 mm2 (Supplementary Figure S2).