Original article

Endotracheal tube cuff pressure assessment: expectations versus reality

Valentyn Sadovyi
1
,
Iurii Kuchyn
2
,
Kateryna Bielka
2, 3
,
Vasyl Horoshko
2, 4
,
Dmytro Sazhyn
2, 3
,
Liubov Sokolova
4

  1. University Clinic of Bogomolets National Medical University, Kyiv, Ukraine
  2. Department of Surgery, Anesthesiology and Intensive Care of the Bogomolets National Medical University, Kyiv, Ukraine
  3. Kyiv Maternity Hospital No. 5, Kyiv, Ukraine
  4. National Military Medical Clinical Center “Main Military Clinical Hospital”, Kyiv, Ukraine
Anaesthesiol Intensive Ther 2024; 56, 4: 241–245
Online publish date: 2024/11/29
Article file
- Endotracheal tube.pdf  [0.11 MB]
Get citation
 
PlumX metrics:
 

Tracheal intubation is one of the most common anesthetic procedures, and, like any other intervention, is accompanied by certain risks. Traumatic intubation is associated with many different complications, ranging from dental damage and post-extubation cough to iatrogenic tracheal injuries [1]. Post-intubation tracheal rupture (PTR) is a rare but serious complication that can be potentially fatal. The incidence and mortality rate of PTR vary according to different studies. Unfortunately, there are no systematic data or extensive meta-analyses on the frequency of these cases [2, 3]. According to various data, the frequency of this complication can range from 1 : 20.000 to 1 : 75.000 during single- lumen tube intubation [4]. And, more often than not, those reports are based either on studies conducted in the 1970s, or individual case reports. The most common clinical signs include subcutaneous emphysema, pneumomediastinum, and unilateral or bilateral pneumothorax. Sometimes PTR can occur without significant visible manifestations, which in turn makes diagnosis difficult [5].

The risk factors of this complication can be divided into anatomical and mechanical. Anatomical factors include congenital anomalies of the trachea (such as weakness of the tracheal membrane), chronic obstructive pulmonary disease (COPD) and other inflammatory lesions of the tracheobronchial tree, diseases that change the position of the trachea (e.g., enlargement of mediastinal lymph nodes or a tumor), corticosteroid use, age, female sex, and short stature. Unsatisfactory results of a physical assessment of the respiratory tract are also separately classified as anatomical factors. The LEMON scale is also prognostically valuable in this context [4, 6]. Mechanical factors include multiple attempts at intubation, inexperience of the anesthesiologist, protrusion of the guide from the tip of the endotracheal tube (ETT), overinflation of the cuff, incorrect positioning of the ETT, repositioning of the tube without deflation of the cuff, inappropriate size of the ETT, severe coughing and movements of the head and neck during intubation (e.g., due to insufficient depth anesthesia/muscle relaxation) [7, 8].

High pressure in the cuff can cause not only such a threatening condition as PTR, but also other complications: cough, tracheal ischemia, sore throat, recurrent paralysis of the laryngeal nerve, or tracheal stenosis [9]. On the other hand, insufficient pressure in the cuff can lead to dislocation of the tube, ineffective ventilation, as well as microaspiration of secretions from the oropharynx. Therefore, the optimal pressure in the cuff of the tube is considered to be 20–30 cmH2O. This range provides reliable protection against aspiration, adequate ETT fixation, and does not significantly increase the risk of iatrogenic damage to the trachea [911].

There are many techniques for measuring ETT cuff pressure. Perhaps the most widespread technique is palpation of the control cuff, but it is primarily based on subjective feeling and does not always reflect the real pressure in the cuff [12]. This technique is much less reliable than mechanical methods of pressure measurement. The use of digital or analog manometers allows for more accurate measurement of cuff pressure, thereby preventing cuff overinflation and under-inflation. Despite that, in developing countries opportunities for cuff pressure control are limited. Only 3.1% (6/196) of the care providers involved in the survey admitted having ever used a tracheal cuff manometer, while 31.1% knew the recommended tracheal cuff pressure. A nationwide telephone survey of anaesthesia faculty fellows revealed that a tracheal cuff manometer was not available, nor had one ever been used in any of the 13 tertiary hospitals surveyed. The ‘pilot balloon palpation method’ and ‘fixed volume of air from a syringe’ were the most commonly utilized methods of cuff pressure estimation by the care providers, at 64.3% and 28.1%, respectively [13]. Of special note are cuff inflation and pressure control techniques, which do not require additional equipment or consumables, but are much more closely correlated with the recommended pressure range. Such methods include the technique of “minimum leakage” or “minimum occlusive volume”. This method is based on gradually inflating the cuff with the minimal volume of air necessary to achieve tightness in the circuit, which enables acceptable pressure to be achieved without using measuring devices, significantly reducing the potential for complications. At the same time, this technique does not lead to an increase in the cost of anesthesia, which is especially relevant in settings with limited resources.

Cuff pressure measurement is an initial step to eliminate complications of tracheal intubation. The purpose of this study was to assess actual cuff pressure using an analog manometer in intubated patients undergoing surgery. The secondary aim was to compare different cuff inflation techniques.

METHODS

The study was approved by the appropriate Commission on Issues of Bioethical Expertise and Ethics of Scientific Research Bogomolets National Medical University (decision #156 from 21/02/22). Written informed consent was obtained from all patients. Ninety patients, randomly divided into 2 groups, were included in the study conducted in Gynecological Department 5 of the Maternity Hospital in Kyiv, as well as the National Military Medical Clinical Center “Main Military Clinical Hospital”. Patients underwent non-urgent surgical interventions, mainly within the abdominal cavity and pelvic organs. Exclusion criteria were a history of difficult airways or multiple attempts of intubation, and BMI over 35 kg m–2. The method of anesthesia in all cases was general low-flow inhalation anesthesia. An oral or nasal ETT was used: 8.0 mm size for male and 7.0 mm size for female patients (cuff material: medical PCV, cuff type: PHT). A standard 20 mL syringe was used to inflate the cuff with air. No N2O was used during the study. In both groups, tracheal intubation was performed after standard induction of anesthesia and mechanical ventilation was started. The inflation of the cuff in the study group (n = 45) was carried out according to the “minimum leakage” method, after which the pressure indicators were recorded with a manometer. In the control group (n = 45), the cuff was inflated by an anesthesiologist and was assessed by palpation of the cuff balloon. After the pressure was deemed satisfactory by subjective assessment, the pressure was measured using a manometer connected to the cuff. Measurements were made in cmH2O within several minutes after intubation. Continuous variables are presented as a mean and standard deviation, while categorical variables are presented as a median with min-max levels. Categorical data were analyzed using Fisher’s exact test. Both groups were divided into “normal” or “overinflated” pressure levels, for which the odds ratio (OR), its standard error and 95% confidence interval (CI) were calculated. P < 0.05 was considered statistically significant.

RESULTS

A total of 40 women and 50 men, with an aver-age age of 44.3 ± 13.6 years and ASA grade 1–2, were included in the study. The t-value for the sample size was 10.42574 (P < 0.001). The mean BMI was 27.1 ± 6.4 kg m–2. The average pressure in the studied group, where the cuff was inflated using the minimal occlusive volume technique, was 30.4 ± 4.9 cmH2O (2.98 ± 0.48 kPa), while in the control group it was 68.9 ± 23.3 cmH2O (6.75 ± 2.28 kPa) (P < 0.001). The criterion for the degree of inflation of the cuff in the control group was the subjective assessment of pressure by the anesthesiologist. In all cases this assessment was satisfactory; however, this did not prevent overinflation of the cuff, which resulted in the measured pressure being several times higher than the recommended norms. The maximum pressure recorded in the control group was 109 cmH2O (10.68 kPa).

In the control group, in only 2/45 (4.4%) the pressure reached the required safe range (i.e. 20–30 cmH2O), while in the study group there were 23/45 (51.1%) such cases (OR = 22.4, 95% CI [4.84], P < 0.001).

Cases of insufficient pressure (less than 20 cmH2O), which could lead to aspiration and other pulmonary complications, were not found in either of the groups.

DISCUSSION

Maintaining a safe pressure range in the ETT cuff is definitely an important factor in preventing the development of various complications of tracheal intubation [11]. However, certain factors prevent the implementation of such an approach in routine practice. Kampo et al. [12] note in their study that cuff pressure devices remain inaccessible to many anesthesiologists, especially in middle- and low-income countries. Although a large number of digital and analog devices have been available on the market for a long time, they are not available to everyone, particularly in resource-limited settings. According to the authors, they did not observe a significant difference in pressure in the cuff between palpation and measurement with a manometer; on average the difference was above 10 cmH2O (0.98 kPa). But the number of complications in the group of empirical determination reached up to 86%, against 0.8% in the group with controlled pressure. Such discrepancies may be due to different approaches in the training of anesthesiologists, insufficient coverage of the problem, etc. These results are subject to further discussion – the study was conducted in 389 women (obstetrics department of the Tamale Teaching Hospital) who were contraindicated for spinal anesthesia or failed neuraxial blockade. We believe that factors such as rapid sequence induction, as well as circumstances that were contraindications to spinal anesthesia, could generally increase the risk of post-intubation complications.

In the study of Souza et al. [10], where they analyzed 25 cases of intubation and prolonged mechanical ventilation in patients in the intensive care unit, the sample of patients was more homogeneous. At the same time, the authors compared two different methods: manual measurement using a manometer and the “minimum required volume” technique. This technique consists of introducing the smallest necessary volume of air or liquid necessary to prevent leakage in the circuit. This does not always correlate with a safe interval of 20–30 cmH2O (1.96–2.94 kPa), although it is definitely also “empirical” in its essence. As was noted in the results of our study, this technique is significantly more likely to result in a pressure falling within the safe range than techniques that rely on the subjective assessment of the pressure by a specialist. Such alternatives deserve attention, especially as a transitional stage in the implementation of safe anaesthesia standards in low- and middle-income countries.

Despite the differences in the designs of the above studies, their results are predictably similar: the use of measuring devices to determine the pressure in the cuff of the intubation tube leads to a decrease in post-intubation complications. That being said, regular measurement of cuff pressure is not routinely performed in many settings.

The relevance of pressure measurement me-thods in the cuff of intubation tubes is even more important in the context of the rising variety of available ETT models, which may differ in their structure, sizes, materials, methods of application, etc. Various studies prove that both the use of excessive pressure introduced by conventional cuffs and ultra-low pressure in small-volume cuffs can lead to various complications. For example, the latter have an increased risk of developing aspiration pneumonia [13].

Despite the fact that some studies note the indisputable advantage of the low-volume, low-pressure cuff, more recent reviews question this conclusion and call for further discussion. The article by Coelho et al. [13] provides a comprehensive narrative review of the use of airway and respiratory devices in the prevention of ventilator-associated pneumonia (VAP), a significant concern in intensive care units due to its association with increased morbidity, morta-lity, duration of mechanical ventilation, and antibio-tic consumption. The review highlights the critical role of specialized artificial airways and devices in reducing the incidence of VAP. However, it also acknowledges the debate over the efficacy of these interventions and their potential adverse effects. This discussion underscores the multifaceted nature of VAP prevention, emphasizing the importance of a balanced approach that considers both the benefits and limitations of technological solutions in the context of comprehensive patient care strategies.

Key insights from the article suggest that while the deployment of advanced airway management devices can play a role in VAP prevention, their application should be judicious and tailored to individual patient needs. It reinforces the idea that prevention strategies must be part of a broader, integrated care protocol that includes diligent hygiene practices, appropriate antibiotic stewardship, and ongoing assessment of patient response to interventions.

Unlike different volumes, different materials from which the cuffs are made do not seem to have such a strong influence on the development of certain complications. The systematic review and meta-analysis conducted by Saito et al. [15] aimed to evaluate the effectiveness of polyurethane (PU) tracheal tube cuffs in preventing ventilator-associated pneumonia (VAP) compared to conventional polyvinyl chloride (PVC) tube cuffs. After a thorough search of several databases, six studies involving a total of 1226 patients were included in the ana-lysis. The results showed that the use of PU cuffs did not significantly reduce the incidence of VAP as compared to PVC cuffs, with a relative risk (RR) of 0.68 (95% CI: 0.45–1.03), indicating no statistically significant difference. The study also noted significant statistical heterogeneity among the included studies (I2 = 65%). Furthermore, the quality of evidence was rated as “very low” and a trial sequential analysis (TSA) indicated that the actual sample size was only 15.8% of the target sample size, suggesting that more research is needed to definitively confirm the effectiveness of PU cuffs in VAP prevention.

CONCLUSIONS

The actual pressure in the cuff measured using an analog manometer in intubated patients is significantly different from the subjective assessment of anesthesiologists. The use of empirical methods of determining pressure, such as palpation of the ETT cuff balloon, is an unreliable method that does not correlate in its effectiveness with mechanical measurement. The minimal occlusion volume technique may be a viable alternative in cases where mechanical measurement of cuff pressure is not available.

ACKNOWLEDGEMENTS

Assistance with the article

none.

Financial support and sponsorship

none.

Conflicts of interest

none.

Presentation

none.

References

1 

Yu Y, Sambol JT, Shakir HA. Conservative management of large tracheal perforation due to traumatic intubation. Am Surg 2023; 89: 2838-2840. DOI: 10.1177/00031348211061099.

2 

Jesani S, Elkatawy S, Ayad S, El-Feki I, Chapla R, Patel H, et al. A rare yet morbid complication of endotracheal intubation: tracheal rupture. Eur J Case Rep Intern Med 2021; 8: 002722. DOI: 10.12890/2021_002722.

3 

Russo E, Caronia FP, Guggino G. Management of post intubation tracheo-esophageal fistula. How, when and why to face off a potentially fatal injury. Diseases of the Esophagus 2022; 35 (Suppl 2): doac051.506. DOI: 10.1093/dote/doac051.506.

4 

Opanasenko MS, Levanda LI, Kshanovskyi OE. Iatrogenic postintubation ruptures of the trachea. Ukrainian Pulmonology Journal 2019; 1: 76-82.

5 

Passera E, Orlandi R, Calderoni M, Cassina EM, Cioffi U, Gutta-dauro A, et al. Post-intubation iatrogenic tracheobronchial injuries: The state of the art. Front Surg 2023; 10: 1125997. DOI: 10.3389/fsurg.2023.1125997.

6 

Katsantonis NG, Kabagambe EK, Wootten CT, Ely EW, Francis DO, Gelbard A. Height is an independent risk factor for postintubation laryngeal injury: height related to intubation laryngeal injury. Laryngoscope 2019; 128: 2811-2814.

7 

Frost A, Ruszkay N, Steinberg TB, Atkins J, Mirza N. Iatrogenic tracheal injuries: case series and review of the literature. ORL 2021; 83: 123-126.

8 

Oo S, Chia RHX, Li Y, Sampath HK, Ang SBL, Paranjothy S, et al. Bronchial rupture following endobronchial blocker placement: a case report of a rare, unfortunate complication. BMC Anesthesiol 2021; 21: 208. DOI: 10.1186/s12871-021-01430-6.

9 

Turner MA, Feeney M, Deeds LJL. Improving endotracheal cuff inflation pressures: an evidence-based project in a military medical center. AANA Journal 2020; 88: 203-208.

10 

Souza BSSD, Souza TAG, Santos CFD, Chicayban PB, Chicayban LM. Minimal occlusive volume is a safe and effective method for adjusting cuff pressure in mechanically ventilated patients. Fisioterapia e Pesquisa 2022; 29: 380-385. DOI: 10.1590/1809-2950/220130311022en.

11 

Kumar CM, Seet E, Van Zundert TCRV. Measuring endotracheal tube intracuff pressure: no room for complacency. J Clin Monit Comput 2021; 35: 3-10. DOI: 10.1007/s10877-020-00501-2.

12 

Kampo S, Anabah TW, Bayor F, Buunaaim ADB, Esquijarosa Hechavarria M, Osman S, et al. Endotracheal tube cuff pressure measurement techniques: safety and reliability: a randomized comparative study. J Clin Anesth Pain Manag 2022; 6: 316-323. DOI: 10.36959/377/376.

13 

Nwosu ADG, Ossai EN, Onyekwulu FA, Amucheazi AO, Ewah R, Onwuasoigwe O, Akhideno I. Knowledge and practice of tracheal tube cuff pressure monitoring: a multicenter survey of anaesthesia and critical care providers in a developing country. Patient Saf Surg 2022; 16: 4. DOI: 10.1186/s13037-021-00311-8.

14 

Coelho L, Moniz P, Guerreiro G, Póvoa P. Airway and respiratory devices in the prevention of ventilator-associated pneumonia. Medicina (Kaunas) 2023; 59: 199. DOI: 10.3390/medicina59020199.

15 

Saito M, Maruyama K, Mihara T, Hoshijima H, Hirabayashi G, Andoh T. Comparison of polyurethane tracheal tube cuffs and conventional polyvinyl chloride tube cuff for prevention of ventilator-associated pneumonia: a systematic review with meta-analysis. Medicine (Baltimore) 2021; 100: e24906. DOI: 10.1097/MD.0000000000024906.

This is an Open Access journal, all articles are distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0). License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 
Quick links
© 2025 Termedia Sp. z o.o.
Developed by Bentus.