2/2017
vol. 13
Original paper
Femoral artery anatomy-tailored approach in transcatheter aortic valve implantation
Anna Olasinska-Wisniewska
,
Aleksander Araszkiewicz
,
Adv Interv Cardiol 2017; 13, 2 (48): 150–156
Online publish date: 2017/05/30
Get citation
Introduction
Transcatheter aortic valve implantation (TAVI) has recently been recognized as a safe and attractive alternative to surgical aortic valve replacement in patients with severe symptomatic aortic valve stenosis who are inoperable or have high risk of cardiac surgery. The procedure results in a significant reduction in mortality as compared to medical therapy [1]. Based on the PARTNER trial [1], the rate of death from any cause at 1 year was 30.7% with TAVI, as compared with 50.7% with standard therapy (p < 0.001), which gives an extremely low number needed to treat of 5 patients with TAVI to prevent 1 death. Most recent studies report an even lower mortality rate (17–18.3%) [2, 3].
Currently, several access sites for valve delivery and implantation are used including femoral, transaxillary, subclavian, transcarotid, transapical and direct aorta. The transfemoral access is a default approach among patients with suitable iliofemoral anatomy, due to its less invasive nature and its feasibility in the majority of patients [4]. Percutaneous puncture or surgical cutdown may be performed to obtain access to the peripheral vessel. Unlike percutaneous access, the surgical approach enables vessel visualization and optimal selection of the puncture site. Whichever technique is preferred, careful evaluation of the size, patency, tortuosity, and the degree of calcification of the iliofemoral arteries is mandatory to determine the feasibility of the transfemoral approach. Vascular complications remain the most frequent ones and significantly affect the outcomes of TAVI [5–7]. Computed tomography (CT) is a recommended method of imaging, due to its value in prediction of vascular complications [8, 9].
Aim
We report on a single-center experience with TAVI performed via femoral access site.
Material and methods
From September 2010 to September 2015, 152 consecutive patients underwent TAVI in our department. Of them, femoral access was obtained in 137 patients, direct aorta in 6 patients, transcarotid in 1 patient and transapical in 8 patients. Medtronic CoreValve prostheses were used in 107 patients (transfemoral access in 101 and direct aorta in 6 patients), Medtronic CoreValve Evolute in 21 patients (transfemoral access in 20 and transcarotid in 1 patient), Boston Scientific Lotus valve in 14 patients (transfemoral access only) and Symetis Acurate in 8 patients with transapical access. In 2 patients the valve was not implanted.
All 101 patients with Medtronic CoreValve implantation from femoral access were included in this analysis.
Our center introduced the femoral artery anatomy-tailored program in 2013 in order to reduce the rate of access-site complications in TAVI patients. The main goal was to find an optimal vessel morphology for a safe percutaneous approach. The following contrast-enhanced CT findings were considered a high risk for the percutaneous approach:
1. Severe calcifications of the femoral artery at the level of the planned puncture site;
2. Diffuse atherosclerotic disease with a large plaque burden at and above the puncture site;
3. Significantly reduced femoral artery dimensions with diameters smaller than 6 mm and a sheath-to-femoral artery ratio greater than 1.0;
4. The presence of circumferential vessel calcifications in femoral or iliac arteries;
5. Extremely tortuous arteries.
All patients with at least one of the above features were scheduled for the surgical cut-down approach. The study cohort was divided into two subgroups: group A – patients treated between September 2010 and December 2012, before introduction of the tailored approach program (n = 34); and group B – patients treated between January 2013 and September 2015 (n = 67). All patients in group A were treated with percutaneous puncture of the femoral artery. Group B contained 48 patients with acceptable femoral anatomy for percutaneous puncture and 19 patients in whom the surgical approach was chosen. In patients with successful percutaneous access, closure with Prostar XL (Abbott Vascular, USA) was used. Surgical access was performed by the cardiac surgeon in the standard manner under direct visualization via a skin incision and subcutaneous tissue dissection with subsequent femoral artery exposure.
The diagnosis of severe aortic stenosis was made based on routine clinical and echocardiographic criteria. Patients were eligible for TAVI on the basis of the institutional heart team’s decision (interventional cardiologist, echocardiographer and cardiac surgeon).
The pre-procedural evaluation included: coronary angiography; transthoracic echocardiography (TTE) and transesophageal echocardiography (TEE); contrast-enhanced computed tomography with off-line reconstructions to evaluate the aorta, femoral and iliac arteries.
The final decision regarding the way of vascular approach was made based on the results of the CT scan.
All procedures were performed under general anesthesia or deep sedation. The TEE monitoring was used in 95% of cases. A temporary pacemaker was inserted from the jugular vein for rapid pacing and to avoid potential consequences of atrioventricular block.
In patients with the percutaneous approach, the Prostar system was introduced before insertion of the vascular sheath (18 Fr sheath inner diameter). In 81 patients, aortic valve predilatation was performed with an undersized Z-MED II-X balloon (NuMED Inc., USA). Once the prosthesis was correctly positioned, expanded and deployed, contrast injection was performed to assess the presence and degree of paravalvular leak. Control angiography was performed to assess vessel patency and possible bleeding.
After the procedure, patients were monitored for heart rhythm and basic life parameters. One day after an uncomplicated procedure, patients began rehabilitation.
Antiplatelet therapy consisted of aspirin 75 mg daily life-long, and in patients with a recent history of coronary angioplasty or acute coronary syndrome clopidogrel 75 mg per day was added, according to current guidelines [10]. In patients with atrial fibrillation, oral anticoagulants were stopped before the procedure, until the international normalized ratio (INR) dropped below 2.0.
The main clinical study endpoints were minor and major vascular access site complications defined in concordance with the Valve Academic Research Consortium-2 (VARC-2) definitions [11]. Consent was obtained from each patient as approved by our institutional ethics committee.
Statistical analysis
Continuous variables were reported as mean and standard deviation and compared using the unpaired t test or nonparametric Mann-Whitney test. Categorical variables were reported as counts or percentages and compared by Fisher’s exact test. All tests were two-sided. P-values less than 0.05 were considered as statistically significant. Statistical analysis was performed using GraphPad InStat 3 (GraphPad Software, Inc. USA). The Kaplan-Meier method was used to calculate the death rate at follow-up.
In a separate analysis, the effect of the learning curve on the vascular complication rate was assessed by exclusion of the first ten patients and comparison between the remaining subgroups.
Results
A total of 101 TAVI patients (mean age: 79.4 ±6.3, range: 61–91, 48.5% male) with CoreValve implantation from femoral access were included in this analysis. Patients’ baseline characteristics are presented in Table I. The two groups did not differ in most baseline characteristics and cardiovascular risk factors. Arterial hypertension and prior myocardial infarction were more common in group B.
During the study period, the access site complication rate significantly decreased from 12 patients in group A (35.3%) to 5 patients in group B (7.5%) (p = 0.0012). Both minor and major access site complications were more frequent in group A, and the difference was statistically significant (Table II).
In-hospital mortality was 8.8% (n = 3) in group A and 1.5% (n = 1) in group B (p = 0.1). Two patients (one female, one male) died due to retroperitoneal bleeding, one male in the course of severe heart failure (all of them in group A) and one female (group B) due to intraprocedural intraventricular rupture. Two deaths were directly related to access site complications in group A, none in group B (5.9% vs. 0%, p = 0.1). The 30-day mortality rate did not differ significantly between groups (p = 0.1) (Table II).
As of December 2015 the clinical follow-up was available at a median of 46.4 months in group A, and 19.4 months in group B. Kaplan-Meier estimated cumulative incidence of death at 30 months was 23.5% in group A, and 12.1% in group B, log rank p = 0.41 (Figure 1).
Six patients required emergency surgical treatment of the femoral and/or iliac artery due to complications, five of them in group A and one in group B (p = 0.02).
In order to eliminate the impact of the learning curve on the complication rate, in the additional sub-analysis we excluded the first 10 patients in group A, and compared those remaining with group B. The result was similar, favoring the tailored approach (Table II).
Discussion
In all 35 initial TAVI patients in whom the percutaneous femoral access site was used, a relatively high rate of access site complications was observed (35%). After this unfavorable experience, we decided to introduce a femoral artery anatomy-tailored approach program in which the decision on the surgical or percutaneous access depended on the preprocedural CT imaging of the femoral and iliac arteries. Therefore, in 48 patients with favorable femoral anatomy a percutaneous puncture was performed, whereas in 19 patients surgical vascular access was applied.
This new approach resulted in significant reduction in minor (20.6% vs. 6%) and major (14.7% vs. 1.5%) access site complications. This reduction was also maintained after exclusion of the first 10 patients from group A in order to eliminate the influence of the learning curve.
In recent years, the vascular complication rate has significantly decreased, from 8–34% in earlier reports [6, 12–14] to 3–15% in the newest studies [15–17]. Many risk factors have been recognized and several techniques proposed to minimize the risk of access site complications. The wide complication rate range may reflect various TAVI vascular approach and vessel closure techniques, the experience of operators, the lack of initial endpoint definitions, or a shift in patient profile towards the lower-risk.
The size of delivery sheaths is of course one of the major risk factors of vascular access site complications. In our study population only 18-French sheaths (inner size) were used. However, it must be pointed out that the first generation delivery systems included larger sheaths, so the incidence of vascular complications in earlier studies may have therefore been higher. Currently, with the newest Medtronic CoreValve Evolute devices and with the Edwards SAPIEN 3 valve, 14-French sheaths are used.
There are several widely recognized patient-related factors which increase the vascular complication risk, including female gender, a lower body surface area, insulin-treated diabetes mellitus, smaller vessel diameter, and higher sheath-to-femoral artery ratio and sheath-to-external-iliac-artery ratio [6]. In the PARTNER trial [6], in patients with major vascular complications the index procedure was longer, with higher contrast volume used and longer fluoroscopy time. Surgical cutdown was used more frequently and the duration of hospitalization was prolonged. The finding of higher incidence of major vascular complications in patients with surgical cutdown may be surprising, especially when compared to our results. However, selection bias as well as the fact that many sites in the trial were at the beginning of their learning curve might play an important role. In fact surgical cutdown was performed in up to 76.8% of patients, whereas suture-based closure device systems were used in only 24.9% of them [6]. The contemporary practice is completely different, with the vast predominance of percutaneous approach.
Holper et al. [18] found that patients with vascular complications had a higher prevalence of three risk factors: female gender, moderate/severe iliofemoral calcification and baseline peak systolic velocity at the femoral artery in bilateral iliofemoral Doppler ultrasound. According to Reinthaler et al. [9] the presence of circumferential iliofemoral calcifications, as representative of severe peripheral vascular disease, is an important risk factor of vascular complications and an independent predictor of mortality after transfemoral TAVI.
Several methods to reduce the access site complication rate have been proposed including the use of preclosure devices, percutaneous vessel repair before TAVI [19], the “cross-over balloon occlusion technique”, in which an endovascular balloon, brought either via the contralateral femoral or radial artery puncture, is inflated proximally to the puncture site prior to removal of the large sheath, use of pre-operative bilateral femoral arterial Doppler ultrasound, particularly in centers with early experience [18], or full surgical management of artery entry [6].
Access and closure via surgical cutdown was performed in more than 75% of patients in the PARTNER trial [6]. As mentioned, this technique was considered to be more predictable, offering more direct control in case of adverse events, especially in less experienced sites at the beginning of the learning curve. According to Toggweiler [20] a surgical cutdown might be particularly desirable in patients where a high puncture is needed due to extensive calcification, a high femoral bifurcation, obesity, or the presence of a femoral stent/graft.
In our opinion surgical cutdown remains a reasonable option in patients with unfavorable femoral anatomy, which puts them at a high risk of access site complications, while low- and moderate-risk patients may benefit from a fully percutaneous approach. As shown in our study, proper selection of patients enables significant reduction of the complication rate.
Two studies [18, 21] comparing the outcomes of the complete percutaneous versus surgical approach revealed inconclusive results. Holper et al. [18] compared, in randomized fashion, the clinical safety and efficacy of percutaneous versus surgical cutdown in a small group of 30 patients. The study demonstrated no difference in the primary endpoint of VARC-2 major and minor complications (25% in the percutaneous and 29% in the surgical arm). However, on the basis of different case scenarios the authors emphasized that percutaneous and surgical approaches still remain complementary techniques of access and intervention in transfemoral TAVI and that surgical cutdown does not preclude arterial complications.
The multi-center Brazilian TAVI registry [21] of 402 patients revealed that the incidence of combined adverse events, including all-cause mortality, life-threatening bleeding, and/or major vascular complications, was comparable in the percutaneous and the surgical groups at 30 days and at 1 year.
The major limitation of the study is the non-randomized, single-center design with a relatively small number of patients. Although we excluded the initial 10 patients from the sub-analysis, still the influence of the learning curve of the operators cannot be ruled out.
Conclusions
The femoral artery anatomy-tailored vascular access approach significantly reduces the incidence of vascular complications in patients treated with trans-femoral aortic valve implantation.
Conflict of interest
Marek Grygier – proctor for Medtronic, other authors declare no conflict of interest.
References
1. Leon MB, Smith CR, Mack M, et al.; PARTNER Trial Investigators. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med 2010; 363: 1597-607.
2. Ludman PF, Moat N, de Belder MA, et al.; UK TAVI Steering Committee and the National Institute for Cardiovascular Outcomes Research. Transcatheter aortic valve implantation in the United Kingdom: temporal trends, predictors of outcome, and 6-year follow-up: a report from the UK Transcatheter Aortic Valve Implantation (TAVI) Registry, 2007 to 2012. Circulation 2015; 131: 1181-90.
3. Debonnaire P, Fusini L, Wolterbeek R, et al. Value of the “TAVI2-SCORe” versus surgical risk scores for prediction of one year mortality in 511 patients who underwent transcatheter aortic valve implantation. Am J Cardiol 2015; 115: 234-42.
4. Toggweiler S, Leipsic J, Binder RK, et al. Management of vascular access in transcatheter aortic valve replacement: part 1: basic anatomy, imaging, sheaths, wires, and access routes. JACC Cardiovasc Interv 2013; 6: 643-53.
5. Ducrocq G, Francis F, Serfaty JM, et al. Vascular complications of transfemoral aortic valve implantation with the Edwards SAPIEN prosthesis: incidence and impact on outcome. EuroIntervention 2010; 5: 666-672.
6. Généreux P, Webb JG, Svensson LG, et al.; PARTNER Trial Investigators. Vascular complications after transcatheter aortic valve replacement: insights from the PARTNER (Placement of AoRTic TraNscathetER Valve) trial. J Am Coll Cardiol 2012; 60: 1043-52.
7. Rodes-Cabau J, Rodés-Cabau J, Webb JG, et al. Transcatheter aortic valve implantation for the treatment of severe symptomatic aortic stenosis in patients at very high or prohibitive surgical risk: acute and late outcomes of the multicenter Canadian experience. J Am Coll Cardiol 2010; 55: 1080-90.
8. Krishnaswamy A, Parashar A, Agarwal S, et al. Predicting vascular complications during transfemoral transcatheter aortic valve replacement using computed tomography: a novel area-based index. Catheter Cardiovasc Interv 2014; 84: 844-51.
9. Reinthaler M, Aggarwal SK, De Palma R, et al. Predictors of clinical outcome in transfemoral TAVI: Circumferential iliofemoral calcifications and manufacturer-derived recommendations. Anatol J Cardiol 2015; 15: 297-305.
10. Windecker S, Kolh P, Alfonso F, et al. 2014 ESC/EACTS Guidelines on myocardial revascularization: The Task Force on Myocardial Revascularization of the European Society of Cardiology (ESC) and the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J 2014; 35: 2541-619.
11. Kappetein AP, Head SJ, Généreux P, et al.; Valve Academic Research Consortium (VARC)-2. Updated standardized endpoint definitions for transcatheter aortic valve implantation: the Valve Academic Research Consortium-2 consensus document (VARC-2). Eur J Cardiothorac Surg 2012; 42: S45-60.
12. Kahlert P, Al-Rashid F, Weber M, et al. Vascular access site complications after percutaneous transfemoral aortic valve implantation. Herz 2009; 34: 398-408.
13. Webb JG, Altwegg L, Boone RH, et al. Transcatheter aortic valve implantation: impact on clinical and valve-related outcomes. Circulation 2009; 119: 3009-16.
14. Thomas M, Schymik G, Walther T, et al. One-year outcomes of cohort 1 in the Edwards SAPIEN Aortic Bioprosthesis European Outcome (SOURCE) registry: the European registry of transcatheter aortic valve implantation using the Edwards SAPIEN valve. Circulation 2011; 124: 425-33.
15. Gilard M, Eltchaninoff H, Iung B, et al.; FRANCE 2 Investigators. Registry of transcatheter aortic-valve implantation in high-risk patients. N Engl J Med 2012; 366: 1705-15.
16. Dato I, Burzotta F, Trani C, et al. Percutaneous management of vascular access in transfemoral transcatheter aortic valve implantation. World J Cardiol 2014; 6: 836-46.
17. Tchetche D, Dumonteil N, Sauguet A, et al. Thirty-day outcome and vascular complications after transarterial aortic valve implantation using both Edwards Sapien and Medtronic CoreValve bioprostheses in a mixed population. EuroIntervention 2010; 5: 659-65.
18. Holper EM, Kim RJ, Mack M, et al. Randomized trial of surgical cutdown versus percutaneous access in transfemoral TAVR. Catheter Cardiovasc Interv 2014; 83: 457-64.
19. Van Mieghem NM, Nuis RJ, Piazza N, et al. Vascular complications with transcatheter aortic valve implantation using the 18 Fr Medtronic CoreValve System: the Rotterdam experience. EuroIntervention 2010; 5: 673-9.
20. Toggweiler S, Webb JG. Challenges in transcatheter aortic valve implantation. Swiss Med Wkly 2012; 142: w13735.
21. Bernardi FL, Gomes WF, de Brito FS Jr, et al. Surgical cutdown versus percutaneous access in transfemoral transcatheter aortic valve implantation: insights from the Brazilian TAVI registry. Catheter Cardiovasc Interv 2015; 86: 501-5.
Copyright: © 2017 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License ( http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
|
|