eISSN: 1897-4295
ISSN: 1734-9338
Advances in Interventional Cardiology/Postępy w Kardiologii Interwencyjnej
Current issue Archive Manuscripts accepted About the journal Editorial board Abstracting and indexing Subscription Contact Instructions for authors Publication charge Ethical standards and procedures
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank
4/2020
vol. 16
 
Share:
Share:
Original paper

Long-term comparison of everolimus- vs. novolimus-eluting bioresorbable vascular scaffolds in real world patients

Beytullah Cakal
1
,
Sinem Cakal
1
,
Oguz Karaca
1
,
Mehmet Onur Omaygenc
1
,
Filiz Kizilirmak Yilmaz
1
,
Haci Murat Gunes
1
,
Ozgur Ulas Ozcan
1
,
Arzu Yıldırım
1
,
Bilal Boztosun
1

  1. Department of Cardiology, Istanbul Medipol University, Istanbul, Turkey
Adv Interv Cardiol 2020; 16, 4 (62): 391–398
Online publish date: 2020/12/29
Article file
- Long-term comparison.pdf  [0.10 MB]
Get citation
 
 

Summary

The largest amount of evidence for bioresorbable vascular scaffold (BVS) use in clinical practice derives from Absorb trials and registries. Comparison of Absorb BVS with metallic stents resulted in increased rates of target lesion failure and device thrombosis in the Absorb BVS group. We investigated whether all BVS platforms created equal outcomes. At 3-year follow-up novolimus-eluting BVS (Desolve Elixir Medical Corp., Sunnyvale, CA) achieved better outcomes compared with Absorb BVS. The scaffold thrombosis rate is still high with Absorb BVS despite pre- and post- dilatation. No scaffold thrombosis was present with Desolve BVS.

Introduction

Bioresorbable vascular scaffolds (BVS) have represented a revolution in percutaneous coronary interventions (PCI) as an alternative to drug-eluting stents (DES) with the goal of complete resorption to eliminate the risk of adverse events due to permanent presence of the metallic stents [1, 2].

The initial analysis of data with the first commercially available everolimus-eluting BVS (eBVS) (Absorb, Abbott Vascular, Santa Clara, CA, USA) in de novo simple lesions showed non-inferior outcomes to metallic DES in patients with stable coronary artery disease in short-term follow-up [3]. However, long-term outcomes from randomized trials showed a significantly higher 3-year rate of scaffold thrombosis for eBVS [4]. Because of these concerns, on September 2017, the manufacturer stopped production of the first-generation Absorb BVS due to low sales. Due to the disappointing outcomes mentioned above, the implementation of BVS was given a class III indication in clinical practice outside of studies in the current European Society of Cardiology (ESC) guidelines [5].

Early positive results of eBVS encouraged many other companies to develop their own versions of BVS, one of which is the Novolimus-eluting BVS (nBVS) (DESolve, Elixir Medical Corporation, Sunnyvale, California, USA), which received the CE mark in May 2014. Sufficient data evaluating either direct or indirect comparison of these two different BVS are presently not available in everyday clinical practice. Therefore, the purpose of this study was to compare clinical outcomes of the two BVS in a real-world population.

Aim

The purpose of the present study based on single-center registry was to compare the long-term outcomes of real-world patients treated with nBVS or eBVS.

Material and methods

Study design and population

This was an observational, retrospective, single-center study of consecutive patients performed in a high-volume, tertiary PCI center, Istanbul, Turkey with an annual PCI volume of approximately 3000 procedures. The study population consisted of all patients who underwent PCI with at least one eBVS or nBVS implantation for coronary heart disease between March 2013 and February 2018. The nBVS has been available for implantation since May 2015. The decision to treat the lesion with either eBVS or nBVS was left to the operator’s choice. All clinical data at follow-up were collected from hospital visits or telephone consultations for all patients. The study was approved by the ethics board of Istanbul Medipol University, Turkey.

Eligible patients were those with reference vessel diameter (RVD) ≥ 2.50 mm, stentable lesions, stable coronary artery disease, or acute coronary syndrome. All treated lesions consisted of stenosis > 50% in native coronary arteries. Major exclusion criteria were defined as left main coronary artery lesions, saphenous vein graft lesions, presence of lesions requiring stents greater than 4.0 mm and less than 2.5 mm, and patients implanted with both nBVS and eBVS. No restrictions were applied for the number of lesions and vessels treated, lesion length or the number of implanted stents.

Study devices

The eBVS consists of an amorphous matrix of poly-L-lactic acid (PLLA), coated with a layer of a 1 : 1 mixture of poly-D-L-lactic acid and 100 µg/cm2 anti-inflammatory everolimus. eBVS is manufactured using extrusion and laser machining techniques, featuring 157 µm strut thickness and comprising in-phase zig-zag hoops linked with a bridge design. The full dissolution process requires 2–3 years to complete [1].

The nBVS is made from a PLLA-derived polymer backbone similar to that of the eBVS and has a strut thickness of 150 µm with an average vessel area coverage of 30% for a 3.0 mm vessel stent. More than 85% of Novolimus at a 3 µg/mm dose is released over 4 weeks with > 95% loss of molecular mass in the first year. The two most remarkable features of nBVS are intrinsic unique self-correcting deployment properties to improve minor strut malapposition and relatively high elasticity with fracture resistance of the scaffold (e.g. 3.0 mm nBVS up to 20 atm with a 5 mm balloon) [6].

Procedure

The choice of BVS was left to the operator’s discretion and the decision for BVS implantation according the principles of PSP (predilation, appropriate sizing, and post-dilation to high pressures) technique was not mandatory but highly recommended. Generally, pre-dilation was performed with compliant or non-compliant balloons, while for more calcified lesions a scoreflex balloon (Orbus Neich, Hong Kong) Angio Sculpt Scoring Balloon (AngioScore, Inc) was preferred. Post-dilatation was performed with non-compliant balloons with the same size of BVS or 0.25–0.5 mm larger.

Quantitative coronary angiography (QCA) was performed offline using standard techniques with automated edge detection algorithms (CAAS 5.7.1, Pie Medical Imaging, Maastricht, The Netherlands) in the hospital’s angiographic analysis center.

Postprocedural management and follow-up

Patients were treated with clopidogrel (600 mg) or prasugrel (60 mg) or ticagrelor (180 mg) loading doses before PCI followed with daily doses of clopidogrel (75 mg) or prasugrel (10 mg) or ticagrelor (180 mg) for 12 months with daily oral aspirin of 100 mg. Clinical follow-up was achieved by medical visits, when patients could not attend visits, they were called and questioned for symptoms.

Outcomes and definitions

The primary outcomes of the study were the patient-oriented composite events (MACE), which included the composite of cardiac death, target vessel myocardial infarction (TV-MI), and clinically driven target lesion revascularization (TLR). Target vessel revascularization (TVR) was defined as repeat PCI or coronary artery bypass graft (CABG) in the target vessel. Deaths were considered cardiac unless a non-cardiac cause was identified. All components of the composite endpoints, and definite thrombosis, were classified according to the Academic Research Consortium [7].

Statistical analysis

Values are presented as mean ± standard deviation (SD) or median (interquartile range (IQR)) for continuous variable or as counts and percentages for categorical variables. The Mann-Whitney U test or the independent sample t test was used for comparison of continuous data and Fisher’s exact test or the χ2 test was used to compare categorical data. Because of the non-randomized nature of the study, propensity score analysis (Fuzzy Python-based extension for SPSS) was used to adjust for differences in patients’ baseline characteristics including age, sex, diabetes, smoking, hypertension, hypercholesterolaemia, previous MI, acute coronary syndrome, family history of coronary artery disease, left ventricular ejection fraction, previous PCI history, previous CABG history, SYNTAX score, and total stent length. Time-to-event variables are presented as Kaplan-Meier curves. P-values of < 0.05 were considered significant. Analyses were carried out using SPSS for Windows, version 25.0 (IBM Corp., Armonk, N.Y., USA).

Results

Baseline characteristics

Baseline characteristics of the overall and propensity matched cohort are presented in Table I. A total of 140 patients with nBVS (n = 202) and 98 patients with eBVS (n = 135) were found to be eligible for matching (Figure 1). Of these, 98 patients treated with 135 eBVS were matched 1 : 1 with 98 patients treated with 136 nBVS.

Figure 1

Study flowchart

/f/fulltexts/PWKI/42699/PWKI-16-42699-g001_min.jpg
Table I

Baseline clinical characteristics

ParameterBefore propensity matchingAfter propensity matching
nBVS (n = 140)eBVS (n = 98)P-valuenBVS (n = 98)eBVS (n = 98)P-value
Age [years]57 ±9.760 ±11.50.0358 ±1060 ±11.50.17
Male119 (85%)79 (81%)0.3781 (83%)79 (81%)0.71
Smoking history72 (51%)40 (41%)0.1144 (45%)40 (41%)0.56
Hypertension87 (62%)62 (63%)0.8664 (65%)62 (63%)0.77
Hyperlipidemia85 (61%)64 (65%)0.4763 (64%)64 (65%)0.88
Diabetes50 (36%)37 (38%)0.7533 (34%)37 (38%)0.55
Previous MI35 (25%)22 (22%)0.6528 (28%)22 (22%)0.33
Previous CABG5 (3.6%)4 (7%)0.224 (4%)4 (7%)0.35
Family history of CAD47 (34%)36 (37%)0.6128 (29%)36 (37%)0.22
New P2Y12 inhibitor86 (61%)53 (54%)0.2659 (60%)55 (54%)0.39
Left ventricle ejection fraction (%)56 ±756 ±9.50.9756 ±756 ±9.50.9
Heart failure5 (3.6%)5 (5%)0.564 (4%)5 (5%)0.73
Clinical presentation table:
Angina116 (83%)1 (83%)0.9778 (83%)81 (80%)0.58
Acute coronary syndrome24 (17%)17 (17%)20 (17%)17 (20%)
SYNTAX Score19.5 ±718.6 ±6.90.319.1 ±6.718.6 ±6.90.55
Total scaffold length [mma]28 (28–56)28 (23–36)0.15a28 (28–56)28 (23–36)0.26b
Procedural success, n (%)139 (99)96 (98)0.5797 (99)96 (98)1

a Median (interquartile range); bMann-Whitney test. MI – myocardial infarction, CABG – coronary artery bypass grafting, CAD – coronary artery disease.

The proportions of patients with stable angina pectoris (83% vs. 80%), acute coronary syndrome (17% vs. 20%) were comparable in the two matched groups. The SYNTAX score (19.1 ±6.7 vs. 18.6 ±6.9; p = 0.55) and total scaffold length (median: 28 mm, IQR: 28–56; median: 28, IQR 23–36, p = 0.26) did not differ between nBVS and eBVS groups, respectively. There were no differences in terms of procedural success between the two groups (98% vs. 99% in eBVS vs. nBVS groups, respectively). In the nBVS group, one patient suffered side branch occlusion, whereas in eBVS there was one scaffold rupture during postdilatation and the second patient had TIMI 1 flow after overlapping of two scaffolds.

Procedural characteristics and quantitative coronary angiography

Angiographic and procedural characteristics of the overall cohort are summarized in Table II and the matched group is presented in Table III. Longer scaffolds were opted in the nBVS group (median: 28 mm, (IQR) 28–28 for nBVS vs. median 28 mm (IQR) 18–28 mm for eBVS; p = 0.002). Overlapping of two BVS was performed in 32 (23.7%) vs. 32 (23.5%) lesions in eBVS and nBVS groups respectively, whereas overlapping with DES was performed in 30 (22.2%) vs. 34 (25%) (p = 0.67) in eBVS and nBVS groups, respectively. The two groups were well balanced according to the lesion complexity including lesion types, bifurcation, CTO, calcification and in-stent restenosis rate.

Table II

Lesion and procedural characteristics in overall cohort

ParameternBVS (n = 202)eBVS (n = 135)P-value
Vessels treated, n (%):0.73
LAD123 (61)70 (52)0.1
LCx29 (14)22 (16)0.62
RCA50 (25)43 (32)0.15
Access route, n (%):
Femoral51 (37)82 (84)< 0.001
Radial89 (63)16 (16)
QCA analysis:
Stenosis percentage [mm]72.71367.5 ±114.90.001
RVD [mm]3.11 ±0.463.16 ±0.430.25
Pre MLD [mm]0.84 ±0.421.02 ± 0.49< 0.001
Post MLD [mm]2.79 ±0.382.62 ±0.46< 0.001
Acute gain [mm]1.95 ±0.521.59 ±0.6< 0.001
Predilatation balloon size [mm]2.87 ±0.422.88 ±0.450.72
Postdilatation balloon size [mm]3.29 ±0.433.27 ±0.470.66
Predilatation, n (%):100%100%
Scoring balloon20%55%< 0.001
Cutting balloon1%0%1
Scaffold diameter [mm]3.1 ±0.443 ±0.410.37
Scaffold length [mm]a28 (28–28)28 (18–28)0.05b
Number of BVS per lesion1.21 ±0.471.15 ±0.360.24
Scaffold length per lesion [mm]31 ±1328 ±10.40.03
Postdilatation, n (%)199 (99)128 (95)0.1
Type of lesion, n (%):
A/B1135 (66)80 (59)0.16
B2/C67 (34)55 (41)
Bifurcated39 (19.3)33 (24)0.26
Calcified55 (28)40 (30)0.6
CTO7 (3.5)7 (5)0.44
Overlap103 (51)62 (46)0.36
In-stent restenosis2 (1)6 (4.4)0.06

a Median (interquartile range), bMann-Whitney test. LAD – left anterior descending, CX – circumflex, RCA – right coronary artery, RVD – reference vessel diameter, MLD – minimum lumen diameter, BVS – bioresorbable vascular scaffold, CTO – chronic total occlusion.

Table III

Adjusted lesion and procedural characteristics (after propensity score matching)

ParameternBVS (n = 136)eBVS (n = 135)P-value
Vessels treated, n (%):0.73
LAD83 (61)70 (52)0.13
LCx20 (15)22 (16)0.72
RCA33 (24)43 (32)0.16
Access route, n (%):
Femoral35 (36)82 (84)< 0.001
Radial63 (64)16 (16)
QCA analysis:
Stenosis percentage [mm]73 ±13.367.5 ±14.40.001
RVD [mm]3.13 ±0.463.16 ±0.430.45
Pre MLD [mm]0.83 ±0.431.03 ±0.48< 0.001
Post MLD [mm]2.78 ±0.392.62 ±0.460.002
Acute gain [mm]1.95 ±0.551.59 ±0.6< 0.001
Predilatation balloon size [mm]2.87 ±0.42.88 ±0.40.90
Postdilatation balloon size [mm]3.29 ±0.423.27 ±0.470.62
Procedural data:
Predilatation (%)100%100%
Scoring balloon22 (16%)74 (55%)< 0.001
Cutting balloon1 (0.7%)0%1
Stent/scaffold diameter [mm]3.1 ±0.443.1 ±0.410.38
Scaffold length [mm]a28 (28–28)28 (18–28)0.002b
Number of BRS per lesion1.18 ±0.541.15 ±0.360.58
Scaffold length per lesion [mm]30 ±12.228 ±10.50.12
Postdilatation, n (%)135 (99)128 (95)0.04
Type of lesion, n (%):
A/B190 (66)80 (59)0.24
B2/C46 (34)55 (41)0.24
Bifurcated28 (21)33 (24)0.44
Calcified30 (22)40 (30)0.17
CTO4 (3)7 (5)0.35
Overlap66 (49)62 (46)0.67
In-stent restenosis2 (1.5)6 (4.4)0.17

a Median (interquartile range), bMann Whitney test. LAD – left anterior descending, CX – circumflex, RCA – right coronary artery, RVD – reference vessel diameter, MLD – minimum lumen diameter, BVS – biovascular scaffold, CTO – chronic total occlusion.

Predilation was performed in every case, but postdilatation was more often used in the nBVS group (p = 0.04). The acute gain value was significantly higher in nBVS compared to eBVS (1.95 ±0.55 vs. 1.59 ±0.6; p < 0.001). The use of newer P2Y12 inhibitors was also similar (p = 0.39).

Clinical outcomes in the propensity matched cohort

Clinical outcomes in the matched population are presented in Table IV. Mean follow-up time for patients implanted with eBVS was 52.9 ±13.9 months and 36.4 ±8 months for nBVS treated patients (p < 0.001). The 3-year follow-up rate was 90% in eBVS and 82% in the nBVS groups. In the propensity matched cohort, at 3-year follow-up, the estimated cumulative MACE incidence (Figure 2) and its singular components were significantly higher in the eBVS group (MACE: 17.3 vs. 6.1%; p log-rank = 0.02, TLR: 16.3% vs. 5.1; p log-rank = 0.02, TV-MI: 0% vs. 8.2%; p log-rank = 0.004) except for cardiac deaths (1% vs.2%; p log-rank = 0.55, for nBVS vs. eBVS, respectively). Scaffold thrombosis involving both early and very late events was detected in the eBVS group (5.1%), but no case of definite scaffold thrombosis was present in the nBVS group (p log-rank = 0.03). The main cause of one non-cardiac death in the nBVS group was pancreatitis, occurring 23 months later. Moreover, the same patient was dialysis-dependent due to contrast nephropathy after nBVS implantation, whereas non-cardiac deaths were linked to gastric and prostate cancers at 23 and 24 months after eBVS implantation.

Figure 2

Kaplan-Meier survival curves at 3-years follow-up according to the eBVS and nBVS. A – MACE-free, B – TLR-free, C – MI-free, D – ST-free

MACE – major adverse cardiovascular events, TLR – target lesion revascularization, MI – myocardial infarction, ST – scaffold thrombosis, eBVS – everolimus- eluting bioresorbable vascular scaffold, nBVS – novolimus-eluting bioresorbable vascular scaffold.

/f/fulltexts/PWKI/42699/PWKI-16-42699-g002_min.jpg
Table IV

Clinical outcomes at 1-year and 3-year follow-up

ParameternBVS (n = 98)eBVS (n = 98)P-value (log-rank)
Clinical outcome at 1 year:
All-cause death1 (1%)1 (1%)0.99
Cardiac death1 (1%)1 (1%)0.99
TV-MI0 (0%)6 (6.1%)0.01
Definite ST0 (0%)3 (3.1%)0.08
TVR2 (2%)8 (8.2%)0.06
TLR1 (1%)8 (8.2%)0.02
MACE2 (2%)10 (10.2%)0.02
Clinical outcome at 3 year:
All-cause death2 (2%)4 (4.1%)0.43
Cardiac death1 (1%)2 (2%)0.98
TV-MI08 (8.2%)0.004
Definite ST05 (5.1%)0.03
TVR6 (6.1%)16 (16.3%)0.03
TLR5 (5.1%)16 (16.3%)0.02
MACE6 (6.1%)17 (17.3%)0.02

[i] MACE – major adverse cardiac event, TV-MI – target vessel myocardial infarction, TLR – target lesion revascularization, TVR – target vessel revascularization, ST – scaffold thrombosis.

Discussion

The present study is the first to report mid-term comparative analysis of nBVS and eBVS outcomes based on propensity score matching. We found that 3-year incidence of MACE mainly driven by higher TLR and MI as well as early and late scaffold thrombosis were higher for eBVS compared to nBVS. Considering that BVS use leads to increased adverse events predominantly driven by the results of eBVS trials, these concerns might not be attributed to nBVS.

After initial enthusiasm in relation to BVS, worse clinical outcomes in terms of TLR and scaffold thrombosis of eBVS, known as the most comprehensively studied BVS, have hampered the clinical use of other BVS [8]. A meta-analysis of the 4 randomized eBVS trials assigning patients to eBVS (n = 2164) or DES (n = 1225) resulted in higher 3-year rates of target lesion failure in the Absorb BVS group (11.7% vs. 8.1%; p = 0.006) [4]. Therefore, the use of BVS has been restricted to clinical studies [9].

Preliminary data from the 5-year follow-up for the DESolve Nx yielded no definite scaffold thrombosis, 4 cardiac deaths (2.5%), and 5 clinically driven TLRs (4.1%). The MACE rate was 7.4% at 2 years [10]. Relatively low-risk lesion subsets (mean lesion length 11.2 ±3.8 mm, Class B2/C lesion rate was 34%) were present in that study in contrast to the current study. Multicenter, post-marketing evaluation of nBVS (100 patients treated with 109 lesions) demonstrated no cardiac deaths, one definite scaffold thrombosis (1%), one MI (1%) and three device-oriented composite endpoint (DOCEs) (3%) at 1 year follow-up [11].

Presently, there are limited data on nBVS in comparison to the eBVS. Reports regarding optical coherence tomography (OCT) and bench comparison of nBVS with eBVS showed that eBVS has fewer peaks per hoop (6 vs. 9), smaller vessel coverage (“footprint”; 27% vs. 30%), and higher radial strength than the nBVS (1.3 atm vs. 1.1 atm) while nBVS provided smaller residual area stenosis (20.1% vs. 14.9%; p < 0.01) and larger maximum (3.2 mm vs. 3.5 mm; p = 0.01) and minimum (2.7 mm vs. 2.9 mm; p < 0.01) scaffold diameters with no strut fracture, suggesting better expansion of the nBVS [12, 13]. Another bench test comparing Magmaris (Biotronik AG, Bülach, Switzerland), Absorb and DESolve BVS showed that Absorb and Magmaris have larger recoil rates and only DESolve did not fracture during side branch dilatation and mini-kissing balloon postdilatation [14]. In our study, QCA analysis showed larger post-procedure minimum lumen diameter (MLD) in the nBVS group despite similar baseline angiographic RVD measured in both groups. Nevertheless, our study does not address the influence of mechanical properties of BVS on clinical outcomes due to the absence of imaging modalities and randomization.

Although no randomized controlled trials or real life studies have been carried out for comparison of nBVS to eBVS, a single center, retrospective comparative analysis using propensity-score matching between eBVS and nBVS (approximately 85% post-dilation rate in both groups with shorter lesion length of 13 mm) showed similar 1-year outcomes in terms of target lesion failure, cardiac death and scaffold thrombosis (1-year target lesion failure 4.7% vs. 4.5%; p = 0.85, respectively), contrary to positive findings for nBVS in the current study [15].

Another noteworthy finding of the current study was absence of scaffold thrombosis in the nBVS group. Scaffold thrombosis has been considered as a consequence of errors during implantation such as inadequate vessel sizing and malapposition [16]. Very late thrombosis and/or restenosis is triggered by scaffold discontinuity of uncovered scaffold struts with intraluminal scaffold dismantling. Therefore, the mechanism of late stent thrombosis/restenosis could be prevented by maximizing the scaffold dimension, embedding struts into plaque, and avoiding acute malapposition, which was strongly related to adverse event rates [17, 18].

Even though the importance of adherence to PSP is frequently emphasized, surprisingly, post-dilatation rates were about < 50%, and varying PSP protocols were found in previously reported large-scale, randomized clinical trials and cohort studies [19]. In our study, extreme attention paid to the implantation technique by the operators was noted in both groups along with slightly lower post-dilatation rates in the eBVS group (pre-dilatation 100% in both groups, post-dilatation 99% in nBVS vs. 95% in eBVS). Also, despite both groups having similar lesion features, initially, more frequent use of scoring balloons in the eBVS group could be explained by the operator’s concern for inadequate lesion preparation. Naturally, most patients receiving eBVS were included in the early period of the study, while patients with nBVS were treated more recently. Thus, it could be speculated that the adoption of the optimal PSP protocol with growing experience over time might have also influenced the reported outcomes. Nevertheless, alarming rates of device thrombosis with eBVS from real life registries and randomized trials possibly have led to a trend towards more often utilization of nBVS lately.

Puricel et al. reported that in a multicenter registry of 1,305 patients implanted with eBVS, the rate of scaffold thrombosis declined significantly in patients when utilizing the strategy optimized for BVS rather than the DES-oriented implantation strategy for BVS procedures [8]. However, despite the optimal implantation practice, in the COMPARE-ABSORB trial, substantially increased risks of thrombotic events were found in complex de novo target lesions, defined as being at least 28 mm in length, in a small vessel, with a pre-existing total occlusion or a bifurcation [19]. Generally, optimization of the implantation technique is considered to be of paramount importance. We still detected higher MACE rates in the eBVS group despite PSP compliance. Therefore, the issue of whether or not to use the PSP protocol for better outcomes remains controversial.

Relatively greater elasticity of nBVS also contributes to lower stent fractures, together with its self-correction property providing better contact of struts and the vessel wall [20]. Routine aggressive postdilatation, unique structural features, shorter bioresorption time (95% reduction in molecular weight by 1 year, with complete absorption by 2 years) of nBVS might be related to the absence of scaffold thrombosis noted in our study. A previous study from our center also showed no scaffold thrombosis with DESolve BVS [21].

In the randomized trials of the Absorb stent (Absorb Japan, Absorb China, Absorb III) more potent new generation P2Y12 inhibitors (ticagrelor, prasugrel) were used in a rate of 24% and 21% [22]. Approximately, more than half of the patients in both groups were given newer P2Y12 inhibitors, but early and late thrombotic events still occurred in the eBVS group beside very high PSP adherence.

The most relevant limitation of our study is the non-randomized nature, instead being a single center registry. Because of the small sample size, it is likely that it was not able to detect small differences between groups. So caution should be exercised in reaching definite conclusions. Lack of intravascular imaging guidance and systematic angiographic follow-up are other limitations of our study. Similarly, intravascular imaging was used in the minority of the patients in BVS trials [23]. Adequately powered randomized trials are needed to determine whether imaging guidance could reduce adverse events. No criteria were pre-defined to decide whether to implant either an nBVS or an eBVS, and thus selection bias cannot be ruled out. Both groups were treated in slightly different time periods, so the results might be biased by the operator experience and adoption of a better PSP implantation protocol. Nevertheless, our findings were obtained in a high-volume center, in which the operators paid meticulous attention to lesion preparation, BVS sizing, and post-dilatation. There was a difference in the loss at follow-up between groups, resulting in unequal duration of exposure to the risk of events. However, we sought to address this limitation by using Kaplan-Meier methods.

Conclusions

nBVS appears to be associated with good mid-term clinical outcomes compared to eBVS. More evidence is needed to evaluate long-term performance of novolimus-eluting biovascular platforms.

Conflict of interest

The authors declare no conflict of interest.

References

1 

Serruys PW, Garcia-Garcia HM, Onuma Y. From metallic cages to transient bioresorbable scaffolds: change in paradigm of coronary revascularization in the upcoming decade? Eur Heart J 2012; 33: 16-25.

2 

Caiazzo G, Kilic ID, Fabris E, et al. Absorb bioresorbable vascular scaffold: what have we learned after 5 years of clinical experience? Int J Cardiol 2015; 201: 129-36.

3 

Ellis SG, Kereiakes DJ, Metzger DC, et al. Everolimus-eluting bioresorbable scaffolds for coronary artery disease. N Engl J Med 2015; 373: 1905-15.

4 

Ali ZA, Gao R, Kimura T, et al. Three-year outcomes with the absorb bioresorbable scaffold: individual-patient-data meta-analysis from the ABSORB randomized trials. Circulation 2018; 137: 464-79.

5 

Neumann FJ, Sousa-Uva M, Ahlsson A, et al. 2018 ESC/EACTS Guidelines on myocardial revascularization. Eur Heart J 2019; 40: 87-165.

6 

Nef HM, Wiebe J, Foin N, et al. A new novolimus-eluting bioresorbable coronary scaffold: present status and future clinical perspectives. Int J Cardiol 2017; 227: 127-33.

7 

Cutlip DE, Windecker S, Mehran R, et al. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation 2007; 115: 2344-51.

8 

Puricel S, Cuculi F, Weissner M, et al. Bioresorbable coronary scaffold thrombosis: multicenter comprehensive analysis of clinical presentation, mechanisms, and predictors. J Am Coll Cardiol 2016; 67: 921-31.

9 

Ali ZA, Serruys PW, Kimura T, et al. 2-year outcomes with the Absorb bioresorbable scaffold for treatment of coronary artery disease: a systematic review and meta-analysis of seven randomised trials with an individual patient data substudy. Lancet 2017; 390: 760-72.

10 

Abizaid A, Costa RA, Schofer J, et al. Serial Multimodality imaging and 2-Year clinical outcomes of the novel DESolve novolimus-eluting bioresorbable coronary scaffold system for the treatment of single de novo coronary lesions. JACC Cardiovasc Interv 2016; 9: 565-74.

11 

Nef H, Wiebe J. A multicenter post-marketing evaluation of the Elixir DESolve(®) Novolimus-eluting bioresorbable coronary scaffold system: first results from the DESolve PMCF study. Catheter Cardiovasc Interv 2018; 92: 1021-7.

12 

Mattesini A, Boeder N, Valente S, et al. Absorb vs. DESolve: an optical coherence tomography comparison of acute mechanical performances. EuroIntervention 2016; 12: e566-73.

13 

Ormiston JA, Webber B, Ubod B, et al. An independent bench comparison of two bioresorbable drug-eluting coronary scaffolds (Absorb and DESolve) with a durable metallic drug-eluting stent (ML8/Xpedition). EuroIntervention 2015; 11: 60-7.

14 

Barkholt TO, Webber B, Holm NR, et al. Mechanical properties of the drug-eluting bioresorbable magnesium scaffold compared with polymeric scaffolds and a permanent metallic drug-eluting stent. Catheter Cardiovasc Interv 2020; 96: E674-82.

15 

Wiebe J, Dorr O, Ilstad H, et al. Everolimus-versus novolimus-eluting bioresorbable scaffolds for the treatment of coronary artery disease: a matched comparison. JACC Cardiovasc Interv 2017; 10: 477-85.

16 

Ortega-Paz L, Capodanno D, Gori T, et al. Predilation, sizing and post-dilation scoring in patients undergoing everolimus-eluting bioresorbable scaffold implantation for prediction of cardiac adverse events: development and internal validation of the PSP score. EuroIntervention 2017; 12: 2110-7.

17 

Stone GW, Abizaid A, Onuma Y, et al. Effect of technique on outcomes following bioresorbable vascular scaffold implantation: analysis from the ABSORB trials. J Am Coll Cardiol 2017; 70: 2863-74.

18 

Gori T, Weissner M, Gonner S, et al. Characteristics, predictors, and mechanisms of thrombosis in coronary bioresorbable scaffolds: differences between early and late events. JACC Cardiovasc Interv 2017; 10: 2363-71.

19 

Danzi GB, Bernelli C, Cerrato E. Outcomes of optimised implantation technique with bioresorbable scaffolds: a pooled analysis of ABSORB-IV and COMPARE-ABSORB trials. Cardiovasc Revasc Med 20120; 21: 559-61.

20 

Schmidt W, Behrens P, Brandt-Wunderlich C, et al. In vitro performance investigation of bioresorbable scaffolds–standard tests for vascular stents and beyond. Cardiovasc Revasc Med 2016; 17: 375-83.

21 

Cakal B, Cakal S, Karaca O, et al. Outcomes of the novolimus-eluting bioresorbable vascular scaffold in real world clinical practice. Minerva Cardioangiol 2020. DOI: 10.23736/S0026-4725.20.05138-5.

22 

Bil J, Gil RJ. Bioresorbable vascular scaffolds-what does the future bring? J Thorac Dis 2016; 8: E741-5.

23 

Serruys PW, Onuma Y. Dmax for sizing, PSP-1, PSP-2, PSP-3 or OCT guidance: interventionalist’s jargon or indispensable implantation techniques for short-and long-term outcomes of Absorb BRS? EuroIntervention 2017; 12: 2047-56.

Copyright: © 2020 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.