eISSN: 2081-2841
ISSN: 1689-832X
Journal of Contemporary Brachytherapy
Current Issue Archive Supplements Articles in Press Journal Information Aims and Scope Editorial Office Editorial Board Register as Author Register as Reviewer Instructions for Authors Abstracting and indexing Subscription Advertising Information Links
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank

3/2009
vol. 1
 
Share:
Share:
abstract:

Original article
HDR and PDR 192Ir source activity control procedures, as the part of the quality assurance system at Brachytherapy Department of Greater Poland Cancer Centre

Grzegorz Zwierzchowski
,
Barbara Błasiak
,
Patrycja Stefaniak
,
Grzegorz Bielęda

J Contemp Brachyther 2009; 1, 3: 157-162
Online publish date: 2009/10/08
View full text Get citation
 
Purpose: One of the main causes of treatment failures in brachytherapy is incorrect source strength specification in planning system or treatment delivery console. Source strength control is the only scheme to avoid such mistakes. The main aim of this work was to present results of three years of HDR and PDR sources activity control.
Material and methods: Study was based on data from 14 192Ir HDR and PDR sources exchanges. Sources were checked three times: at the exchange day and after one and two months. Measurements were performed twice with thimble chamber (PMMA phantom), and well chamber. The source strength were measured as air – kerma and recalculated to activity.
Results: Source activities measured using well chamber and thimble chamber, as well as activities provided by planning system, were presented for PDR and HDR, respectively. Differences between results obtained using each chamber and activities from planning system were presented graphically. The calculated and measured activities differed less than 5%. Wilcoxon test was performed as well, no statistically significant differences were observed among HDR or PDR activities.
Conclusions: Checking of source parameters is one of the most important parts of quality control system in brachytherapy facilities. Well chamber and thimble chamber based dosimetry systems are fast and reliable tools for 192Ir source parameters checking in working brachytherapy department conditions.
keywords:

quality assurance, source activity check

 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.