Revision of the dosimetric parameters of the CSM11 LDR Cs-137 source

Antonio Otal
,
Juan Manuel Martínez-Fernández
,
Domingo Granero

J Contemp Brachyther 2011; 3, 1: 36-39
Online publish date: 2011/03/31
Article file
Get citation
 
 

Purpose

Although the use of low dose rate (LDR) sources in brachytherapy has declined in recent years in favor of the high dose rate sources (HDR), the use of LDR 137Cs sources is still prevalent in some centers. The reason for this is the necessity of the existence of updated dosimetric data tables that allows proper treatment planning. One of these LDR sources is the CSM11 source (BEBIG GmbH, Germany). The earlier study of this source by Ballester et al. [1] was performed using the Monte Carlo code GEANT3 [2]. Ballester et al. obtained dosimetric data for this source with statistical uncertainty too elevated (see figures 2 to 5 of the publication) for the recommendations given in the TG-43U1 protocol [3]. Therefore, it is necessary to update the data to reduce the noise. The purpose of this study was to obtain the dosimetric data for the CSM11 source without noise using the Monte Carlo code GEANT4 [4] in order to improve the existing data.

Material and methods

The dimensions and materials of the LDR 137Cs CSM11 source model (BEBIG GmbH, Germany) were obtained from the study by Ballester et al. (Fig. 1). The active part was a seed of pollucite (Cs2O, Al2O3, 4SiO2,  = 2.9 g/cm3) encapsulated in a stainless steel (Z3 CDN 18/12). It is important to note that in this study the origin of coordinates was situated at the middle of the source active part, not in the center of the source as for other 137Cs LDR sources. In Ballester et al. study, the origin was located in the geometric center of the source.

The Monte Carlo code GEANT4 (version 9.2) was used to obtain the dose rate distributions around the CSM11 source. The 137Cs photon spectrum applied in the Monte Carlo simulation was obtained from the Nudat database [5]. Physical models used for Compton scattering, photoelectric effect and Rayleigh scattering processes were acquired from the low-energy package of GEANT4 which uses the cross sections from the EPDL97 library [6] as recommended in TG-43U1. In the simulations the kerma track-length estimator [7] was used to optimize the simulation time and to reduce the statistical noise of the data. Kerma was calculated assuming electronic equilibrium in the region of interest [8] and thus in that region kerma coincides approximately with dose. The cutoff energy for photons and electrons was 10 keV.

The CSM11 source was placed at the center of a spherical liquid water phantom 40 cm in radius which was acting as an unbounded phantom up to 20 cm from the source center [9,10] in order to obtain the dose rate distribution around the CSM11 source. A grid system composed of 400 × 180 concentric spheres sections of 0.05 cm thick with an angular size of 1 was used to score kerma. To acquire sK the source was surrounded by vacuum except for a small cylindrical air cell of 0.1 cm in diameter and air kerma was scored in the air cell. The composition by weight and density of water and air used for the simulations was that recommended in the TG-43U1 update.

Results

The results obtained by the Monte Carlo code were analyzed with the ROOT data analysis framework [11] enable to attain the along-away Cartesian 2D lookup data and the TG-43 dosimetric parameters. The along-away 2D Cartesian lookup data D.(y,z) is presented in Table 1. The TG-43 parameters were obtained through the data D.(r,) using the geometric factor GL(r,θ) with the line-source approximation. The active length of the source was L = 0.32 cm. We found a value for the dose rate constant of Λ = 1.09 ± 0.002 cGy h-1 U-1. The anisotropy function obtained, F(r,), is presented in Table 2. The radial dose function gL(r), is shown in Table 3 compared with that of Ballester et al. As could be observed in Table 3 both datasets are coincident within statistical uncertainties.

Discussion

Following the recommendations of the TG-43U1 report, dose rate uncertainties were evaluated considering the type A or statistical uncertainty contributions inherent to the Monte Carlo technique and the type B contributions arising from uncertainties in the cross section databases in the geometry of the source in the photon spectra and in other possible systematic uncertainties. In case of type A uncertainties, we distinguished between two intervals. In case where  > 10 and  < 170°, the obtained values for the uncertainty below 0.2% and close to the longitudinal source axis for  < 10° and  > 170°, where the statistical uncertainty was less than 1%. The statistical uncertainty in the case of the simulation of air-kerma strength was less than 0.1% (type A uncertainty). The combination of both statistical uncertainties gives type A uncertainties of 0.2% for the first interval and 1% for the points located near the longitudinal source axis. The uncertainties due to the source construction processes were not evaluated in this study. The uncertainty contribution associated to the cross section libraries and other systematic uncertainties associated with the MC calculation were not evaluated, however a conservative value of 1% seems appropriate [12]. The final dose rate values, combining both types A and B uncertainties, for points near the longitudinal source axis were less than 1.4% and less than 1% for the other points.

Conclusions

In this study the dosimetric parameters of the CSM11 source were acquired in order to obtain the TG-43 parameters of the source. The results were coincident within the statistical uncertainties with that obtained for the same source by Ballester et al. This study presents insignificant statistical noise in contrast with the study of Ballester et al., thus the new dosimetric data for the CSM11 source fulfils the requirements of the TG-43U1 report to be used in clinical practice as a reference data and improves the existing dosimetric data for this source.

References

 1. Ballester F, Lluch JL, Limami Y et al. A Monte Carlo investigation of the dosimetric characteristics of the CSM11 137Cs source from CIS. Med Phys 2000; 27: 2182-2189.  

2. Brun R, Bruyant F, Maire M et al. GEANT3, CERN DD/EE/ 84-1, 1987.  

3. Rivard MJ, Coursey BM, DeWerd LA et al. Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med Phys 2004; 31: 633-674.  

4. Agostinelli S, Allison J, Amako K et al. GEANT4 – A Simulation Toolkit. Nuc Ins Meth A 2003; 506: 250–303. See also http://geant4.web.cern.ch/geant4, last accessed 28 December 2010.  

5. National Nuclear Data Center. Nuclear data from NuDat, a web-based database maintained by the National Nuclear Data Center, Brookhaven National Laboratory, Upton, NY; http://www.nndl.bnl.gov/nudat2/.  

6. Cullen D, Hubbell J, Kissel L. EDPLD97: The evaluated Photon Data Library ’97 version. Lawrence Livermore National Laboratory, UCRL-50400 1997: Vol. 6 Rev. 5.  

7. Williamson JF. Monte Carlo evaluation of kerma at a point for photon transport problems. Med Phys 1987; 14: 567-576.  

8. Ballester F, Granero D, Pérez-Calatayud J et al. Evaluation of electronic equilibrium and electron contribution to dose near brachytherapy sources. Med Phys 2009; 36: 4250-4256.  

9. Melhus CS, Rivard MJ. Approaches to calculating AAPM TG-43 brachytherapy dosimetry parameters for 137Cs, 125I, 192Ir, 103Pd, and 169Yb sources. Med Phys 2006; 33: 1729-1737.

10. Pérez-Calatayud J, Granero D, Ballester F. Phantom size in brachytherapy source dosimetric studies. Med Phys 2004; 31: 2075-2081.

11. Brun R, Rademakers F. ROOT – an object oriented data analysis framework. Nucl Instrum Methods Phys Res A 1997; 389: 81-86. See also http://root.cern.ch last accessed 28 December 2010.

12. Granero D, Vijande J, Ballester F et al. Dosimetry revisited for the HDR 192Ir brachytherapy source model mHDR-v2. Med Phys 2011; 38: 487-494.
Copyright: © 2011 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 