The present report describes a novel approach to in-stent thrombosis in the left pulmonary artery treated with the Rotarex thrombectomy system in patients with single ventricle anatomy after cavo-pulmonary (Glenn) shunt.
We report a case of a 13-year-old girl with single ventricle anatomy (a double outlet right ventricle with a hypoplastic left ventricle) after cavo-pulmonary (Glenn) shunt, who was referred to our hospital with deteriorating exercise capacity and low arterial oxygen saturation (75%).
Cardiac catheterization was performed and angiography demonstrated left pulmonary artery stenosis with significant antegrade pulmonary flow from a single ventricle (Figure 1 A). A 12 Fr Mullins long sheath was introduced and a covered stent (CP 8ZIG 45 mm, NuMED) mounted on a 14 mm diameter BiB balloon (NuMED) was implanted in the middle portion of the left pulmonary artery, and thus the antegrade pulmonary flow was closed and the pulmonary stenosis was resolved (Figure 1 B). Standard anti-platelet therapy with acetylsalicylic acid was introduced and the patient was discharged 6 days later. The patient was considered as a candidate for a Fontan operation and the next cardiac catheterization was planned in 6 months.
Three months later the patient was admitted to our hospital again with signs of progressive cyanosis (arterial oxygen saturation: 65%).
Urgent cardiac catheterization was performed and angiography demonstrated an occlusion of the previously implanted stent in the left pulmonary artery (Figure 1 C). Conservative medical management with heparin was administered. After 1 week of therapy, no clinical evidence of improvement had occurred. Because the patient was considered a poor surgical candidate due to numerous previous sternotomies, the decision to recanalize the left pulmonary artery with the Rotarex (Straub Medical) thrombectomy system was made. The occluded vessel was crossed through with a 0.014′′ Asato guidewire (Asahi Intecc) and a 4 Fr vertebral catheter. Multiple passes through the thrombus were made with different catheters and guidewires; thus partial mechanical fragmentation of the thrombus was performed. An 8 Fr Rotarex catheter was advanced and multiple blood clots were aspirated and removed. Subsequently, an 18-mm diameter Atlas Gold balloon (Bard Medical) was inserted and inflated. Control angiography showed complete recanalization of the left pulmonary artery, but inside its branches thrombi were still present. For this reason, numerous aspirations of this material were made, resulting in a large amount of embolic material. In subsequent angiography, the flow to all segments of the left lung was confirmed (Figure 1 D). After the procedure, a blood transfusion was required. Standard anticoagulation and antiplatelet treatments were initiated. In the following days, we observed slightly higher arterial oxygen blood saturation of 75–80% and better exercise capacity.
Pulmonary artery stenosis associated with a significant anterograde pulmonary flow is rarely found in patients with single ventricle anatomy after the Glenn operation. Covered stent implantation resolves both problems and this procedure has been recently described [1]. Nevertheless, a nonpulsative, slow pulmonary blood flow, a high hematocrit level and PTFE covering of the stent create favorable circumstances for in-stent thrombosis.
A conservative therapeutic strategy for in-stent thrombosis includes anticoagulant or thrombolytic therapy. If pharmacological reperfusion therapy has failed, surgical embolectomy should be deliberated. Although pulmonary embolectomy is considered a technically simple operation, patients with numerous previous sternotomies appear to have a worse outcome. Alternatively, catheter-based mechanical thrombectomy is one of the treatment options. However, it has been rarely reported in patients with single ventricle anatomy [2, 3].
Hirono reported two cases of percutaneous mechanical thrombectomy for in-stent thrombosis in the pulmonary artery in patients after Fontan surgery [2]. In both cases, the mechanical thrombectomy system was used for the treatment of a recent thrombus, which formed between ten days and one month after stent implantation. Only in the first case did the pulmonary artery remain patent. The time from the beginning of clot formation to intervention may be crucial. Even 2 weeks after the incident, the thrombus may be partly converted to fibrous connective tissue. In these circumstances, anticoagulation and thrombolysis therapy are ineffective. Also, mechanical thrombectomy shows the best effectiveness in a newly formed thrombus; nevertheless, we achieved a satisfactory effect in patients 3 months after stent implantation.
One should pay attention to the fact that the Rotarex thrombectomy system is generally intended for use in peripheral arteries due to the extraordinary suction power of this device [4]. In our opinion, this device is both safe and effective in pulmonary arteries when treating occlusion in a previously implanted stent.
In conclusion, mechanical thrombectomy for in-stent thrombosis in children constitutes an alternative option when conservative treatment is ineffective and surgery is contraindicated.