Purpose
Endometrial and cervical cancer are two of the most common gynecological malignancies that afflict women worldwide [1]. Vaginal cuff brachytherapy (VBT) has become an essential technology used for adjuvant treatment of endometrial cancer of patients who are at high-risk of recurrence post-hysterectomy [2, 3] and carefully selected post-hysterectomy patients with cervical cancers [4]. The selection of an appropriate VBT applicator is important to achieve optimal results.
Single-channel cylinder (SCC) is the most common applicator for VBT [5]. However, it is limited in its’ ability to sculpt clinical target volume (CTV) and organs at risk (OARs) doses due to its’ single-source anisotropy. On the contrary, multichannel cylinder (MCC) applicator presents with more peripheral channels to allow for dose flexibility [6]. It is reported that MCC significantly reduces doses to the rectum (by approximately 15%) and bladder (by 1.4-14%) compared with SCC [7-9]. At the Peking Union Medical College Hospital, we have used in-house MCC applicator for adjuvant VBT since 2003, and have observed a 5-year overall survival and local recurrence rates of 96.3% and 2.9%, respectively, in patients with stage I high-intermediate-risk endometrial cancer [10]. However, post-hysterectomy vaginal morphology varies due to tumor’s location, surgical approach, and surgical closure, thus cylindrical applicators cannot fit all the patients. Customized vaginal mould can adapt and show the shape of the vagina, and it can be used for irregular vaginal vault configuration [11-14].
Three-dimensional (3D) printing has enabled a physical creation of customized vaginal moulds with optimized catheters embedded in applicator. Wiebe et al. found that a 3D-printed applicator could offer favorable dose distribution at the vaginal apex for a ‘dog ear’ vaginal configuration, with sufficient and effective dwell positions [15]. However, previous studies investigating the use of this individual applicator in vaginal brachytherapy have focused on interstitial brachytherapy [16-22], whereas reports on vaginal intracavitary brachytherapy are scarce [15, 23]. As such, whether 3D-printed applicators are superior to MCCs in clinical practice remains unknown.
We adopted 3D-printed individualized applicators at our institution in 2015. Under a hypothesis that 3D-printed applicators would provide better CTV coverage and OARs’ protection, we performed this study to compare the dosimetric parameters of 3D-printed plans for high-dose-rate (HDR) VBT to those of MCC plans at a given dose. Additionally, we explored and classified post-hysterectomy vaginal morphologies via analysis of the 3D-printed applicator shapes.
Material and methods
Patients’ selection
We extracted consecutive patients with gynecological cancers who required VBT after hysterectomy, and those with lesions thicker than 10 mm were excluded. In total, 48 patients were analyzed including 26 with endometrial cancer, 16 with cervical cancer, 3 with recurrent endometrial cancer, and 3 with recurrent cervical cancer.
Applicators
The in-house multichannel cylinder applicator (Patent No. ZL201320564893.3) comprised one central channel and six peripheral channels; its’ details have been described previously [10]. The development of 3D-printed applicator is shown in Figure 1. There were three critical steps involved in generating a 3D-printed applicator: 1. Vaginal packing with gauze, in which the vagina was filled with narrow, thin gauze pre-soaked in diluted diatrizoate meglumine. Then, computed tomography (CT) images were obtained and the gauze was withdrawn, which took about 10 minutes; 2. Extracting the gauze and placing source catheters, in which CT images were transferred to 3D Slicer software (http://www.slicer.org) to determine the volume of vagina gauze, served as a prototype of a 3D-printed applicator. The outer catheters were then placed approximately 5 mm away from the applicator surface, with 10 mm space between each other. It took about 30 minutes for image extraction and model design; 3. Production and quality assurance of 3D-printed applicator, in which individualized applicator was made from a biocompatible OBJET MED610 polymer (Stratasys Ltd., Rehovot, Israel), using Eden 260VS 3D printer for 1-2 hours. The applicator was then subjected to multiple physical evaluations to ensure its’ suitability for clinical use. Additionally, we classified the 3D-printed applicators depending on their apex shapes.
CT simulation and treatment plans
All the patients received pelvic CT simulation with 3 mm slice thickness to confirm the position of applicators. Contouring and treatment planning were developed using Oncentra (Elekta, Stockholm, Sweden) brachytherapy treatment planning system.
Dosimetric parameters
The goal was to cover 90% of CTV using the prescribed dose of 500 cGy at 5 mm depth. CTV was defined as a 5 mm expansion beyond the vaginal applicator surface; any surrounding air gaps and lesions beyond 5 mm would be considered when contouring CTV, while OARs were excluded from CTV. Treated length was approximately 1/3 to 1/2 of the vagina, depending on pathological sub-type, International Federation of Gynecology and Obstetrics stage, and risk of recurrence.
CTV-1 cm was defined as proximal 1 cm of CTV from the vaginal apex, which normally represented the true vaginal cuff volume [6]. Measuring this parameter was reasonable, given that the most common location of post-operative endometrial cancer recurrence is the vaginal vault [24, 25]. Four quality indices were evaluated, including conformity index (CI), which was defined as a ratio of volume receiving ≥ 95% of the prescription dose to total CTV (CI = V95/VCTV); dose homogeneity index (DHI), which was defined as a ratio of volume receiving 100-150% of the prescription dose to total CTV (DHI = V100-150/VCTV); dose non-uniformity rate (DNR), defined as a ratio of volume receiving at least 150% of the prescription dose to that receiving at least 100% of the prescription dose (DNR = V150/V100); and overdose index (OI), defined as a ratio of volume receiving at least 200% of the prescription dose to that receiving at least 100% of this dose (OI = V200/V100). Overall, the ideal CI value was 1, and a higher DHI indicated a more homogeneous dose distribution in the treatment volume, while a smaller DNR showed a lower non-uniform dose distribution in the treatment volume. The ideal OI rate was 0.
Wilcoxon signed-rank test was applied to compare differences between MCC and 3D-printed applicators. Fisher’s exact probability method was used to compare the incidences of air gaps associated with different vaginal morphologies on CT images. A two-sided p-value < 0.05 was considered statistically significant.
Results
Table 1 shows the patients’ characteristics. Air gaps were present in 24 of the 48 MCC CT simulation sets; the median air gap volume was 0.11 cm3 (range, 0.02-0.33 cm3). Post-hysterectomy vaginal morphologies were classified into 5 types based on 3D-printed applicator shapes (Figure 2), including dome-column, Gothic arch-column, two dog ears-column, one dog ear-column, and ‘up wide and low narrow’.
Table 1
Table 2 shows the comparison between dosimetric parameters of the MCC and 3D-printed plans. In terms of CTV coverage, V100 was slightly larger in the 3D-printed plan than in the MCC plan (90.02% vs. 90.01%, p < 0.001) and D98 in the 3D-printed group (429.11 cGy) was higher than that in the MCC group (400.08 cGy, p < 0.001). All four quality indices (CI, DHI, DNR, and OI) were superior in the 3D-printed plans than in the MCC plans. When evaluating CTV-1 cm, the difference in V100 was significant and favored the 3D-printed plans (81.85% vs. 72.47%, p < 0.001). Additionally, mean D90 was significantly higher in the 3D-printed plans than in the MCC plans (470.05 vs. 422.11 cGy, p < 0.001), indicating inadequate vaginal apex coverage by both applicators, especially during MCC treatment.
Table 2
Table 3 illustrates the dosimetric comparison for OARs between the two plans. D2cc and D0.1cc of the bladder and D0.1cc of the rectum were lower in the 3D-printed plan than in the MCC plan. No differences were noted between the two plans in terms of D2cc of the rectum and D2cc and D0.1cc of the small intestine. However, D2cc and D0.1cc of the sigmoid in the 3D-printed group were higher than those in the MCC group (129.34 vs. 99.55 cGy, p < 0.001; 184.48 vs. 154.25 cGy, p = 0.001, respectively).
Table 3
Table 4 shows the air gap incidences associated with the five vaginal morphologies. The two dog ears-column had the highest air gap rate at 83.3%, followed by the Gothic arch-column (75%), one dog ear-column (40%), dome-column (33.3%), and ‘up wide and low narrow’ (25%). The differences between these rates were not statistically significant.
Discussion
Our study was the first to perform dosimetric comparison between MCC and 3D-printed applicators for HDR VBT. Moreover, this was the first study to differentiate post-hysterectomy vaginal morphologies into 5 types via analysis of 3D-printed applicator shapes.
An ideal applicator should perfectly fit the shape of vaginal cavity and provide enough dwell positions to avoid both target underdosage and OARs overdosage. In terms of cylindrical applicators used during VBT treatment, commercial SCCs and MCCs have the same shape and size; however, the latter have more catheters that enable flexible dwell positions. MCC applicators have become significantly more popular in recent years [5, 26]; however, the vaginal anatomy is non-uniformly cylindrical. Air gaps are often observed during VBT treatment when using cylindrical applicators. One of the most central features of brachytherapy is the inverse square factor (dose rate = K/distance2). A displacement of 1-2 mm of vaginal wall caused by an air gap can decrease the dose to vaginal mucosa by more than 10% [27]. Another study has shown that a 1 mm displacement of vaginal mucosa would result in a mean dose reduction of 7.3% and 9.2% at a 5 mm depth and at vaginal surface, respectively [28]. Choo et al. has found that more than 90% of lymphatics lie within 2-3 mm of stretched vaginal mucosal surface [29]. Therefore, HDR-VBT was prescribed at a 5 mm depth from the applicator surface to allow for a 2 mm tolerance [30].
If air gaps larger than 2 mm cannot be reduced, customized applicators may provide a solution. Magné et al. reported on a vaginal mould applicator, which can allow for tailored treatment in terms of morphologic optimization [12]. There were some procedures for this vaginal mould: 1. A vaginal impression was made to outline patient’s anatomy and tumor topography; 2. A rough hollow acrylic applicator was made; and 3. The position of catheters was confirmed to provide satisfactory dosimetric coverage. The effect of vaginal impression made from an alginate material is similar to that of our vagina packing with thin strips of gauze. Handling of chemicals in the vaginal mould-making process was not easy for the staff, but 3D-printed applicator did not have this problem. Similar to MCC, molded applicator flattened the irregular vaginal apex by expanding the vaginal wall [13]. It may be uncomfortable and difficult for vaginal mould to flatten the long “dog-ear” configuration (Figure 1D), but easy for the 3D-printed applicator to adapt to this configuration without discomfort. In addition, for some ‘up wide and low narrow’ configuration patients, it may be difficult to extract the vaginal impression from the vagina once it has hardened, but easy to remove the thin gauze strips one by one. Furthermore, the 3D-printed applicator can be designed with two parts to be inserted into the narrow vaginal introitus with minimal patient’s discomfort. Wiebe et al. also reported similar two-part design [15]. Additionally, the arrangement of catheters should be of special interest. Our 3D-printed applicator had a circle of peripheral catheters and several central catheters; therefore, there were more catheters than 3 in Magné et al.’s, and more flexible arrangement than that in Wiebe et al.’s linear arrangement. Furthermore, for patients with obvious residual lesions before VBT, the catheters can be preliminarily designed by radiation oncologists and physicists based on CT images with vaginal gauze. Then, target coverage and dose distribution can be roughly evaluated before the 3D-printed applicator is made.
Our results confirmed that 3D-printed applicators could cover a larger CTV than MCC applicators, especially CTV-1 cm. Since the optimization condition was limited to D90 = 500 cGy, the improvement in CTV V100 was only 0.01%. However, the 3D-printed applicators offered higher D98 and more homogeneous dose distributions within CTV. Additionally, our study showed significant improvements in CTV-1 cm V100 (by 13%) and D90 (by 11%) when compared with MCC. When comparing the 3D-printed applicators to standard SCC counterparts in patients with ‘dog ear’ vaginal configuration, Wiebe et al. found that CTV V100 improved by 13.2%, whereas the dose of ‘two ears’ area increased by 113%, with more uniformity (V200 reduction of 6.8%). Simultaneously, total doses to OARs when using 3D-printed applicators were lower than those observed when using SCCs [15]. In our study, the 3D-printed applicators showed superiority over MCCs in terms of the bladder and rectum protection. However, we found it would cause higher undesired doses to the sigmoid. It is possible that better coverage of CTV-1 cm was achieved at the expense of a higher dose to the sigmoid.
Although vaginal anatomy was usually considered as a cylinder during VBT treatment, it has been found to resemble a funnel [31]. From the vaginal introitus to the cervix, the distended vagina was consistently found to have three zones: a superficial sphincteric narrow zone, a central wedge-shaped transition zone, and an expanded forniceal zone. The narrow introitus limited the size of applicator; hence, air gaps were likely to occur at the expanded fornix when using cylindrical applicators with a suitable diameter for the vagina introitus. Even if no air gap was observed on CT image, the vaginal folds at the fornix were often piled up, which goes against the principle of stretching the vaginal mucosa as much as possible during VBT treatment. The 3D-printed applicator can conformally fit into the three consistent zones and extend the vaginal folds without discomfort. Various surgical approaches may deviate from the vaginal axis and affect the angle between the upper and lower vaginal segments [32], thereby affecting the shape of the vaginal cuff [33, 34]. Park et al. found that vaginas demonstrated an elliptical rather than circular cross-sectional anatomy [9]. As the configuration of vagina is complex, we failed to find quantitative indicators for a classification of 3D-printed applicators, and roughly divided them into five types to show vaginas’ morphologies.
The dome-column (43.8%) was the most commonly observed vaginal morphology post-hysterectomy in this study, and the best-suited for cylinder applicators. The Gothic arch-column, two dog ears-column, and one dog ear-column configurations of the vaginas, which presented with high air gap rates when using MCC, may preferably use a 3D-printed applicator. We also found that it was clinically feasible to judge the vaginal morphology through gynecological examinations before CT simulation. For example, it was easy to differentiate two dog ears-column configurations in gynecological examinations, indicating that those patients may need a 3D-printed applicator. However, no significant differences were found in air gap incidences among the five vaginal morphologies, partially owing to the small sample size.
The thickness of lesions treated in this study was limited to 10 mm. For patients with tumor sizes > 10 mm [18], paravaginal extension or pararectal space involvement [16], and insufficient target coverage [20], combined intracavitary/interstitial techniques should be used. Novel 3D-printed intracavitary applicators based on intensity-modulated techniques has been shown to be clinically feasible for treating vaginal tumors thicker than 10 mm [35]. These applicators can potentially improve therapeutic ratio by facilitating OARs’ dose reduction, CTV dose escalation, or both [36].
There were limitations in this study. Our in-house MCCs only had a single-fixed diameter of 26 mm; however, this was suitable for most patients treated at our center. Furthermore, the incidence of air gaps on MCC CT images was 50%, which was lower than that reported by Cameron et al. [37] and Richardson et al. [38], but similar to Hassouna et al.’s reported rate of 58% [27], which indicated good contact between in-house MCC and vagina. Secondly, the relatively small sample size confounded the identification of vaginal morphology most likely to generate air gaps. Thirdly, it was a retrospective study.
Conclusions
Compared to MCC, the 3D-printed applicators delivered a higher dose and larger volume, and offered more homogeneous and conformal target coverage. Further research on vaginal morphology would be valuable to help screen out patients who require 3D-printed applicators, and also to provide guidance for a design of new applicators.