Original paper Permanent prostate brachytherapy extracapsular radiation dose distributions: analysis of a multi-institutional database
J Contemp Brachytherapy 2013; 5, 3: 117–121
Online publish date: 2013/10/02
Get citation
PurposePermanent prostate brachytherapy represents a highly efficacious treatment for clinically localized prostate cancer with a defined relationship between dosimetric quality, biochemical outcome and complications [1,2]. Previous studies have demonstrated that long-term cancer control is related to intra-prostatic radiation dose and periprostatic treatment margins [3-6]. However, treatment margins can vary markedly between patients with comparable intra-prostatic dose distributions [5]. Because brachytherapy dose decreases by as much as 20 Gy per millimeter at the periphery of the target volume, as many as 50% of patients with a pre-treatment prostate specific antigen (PSA) < 10 ng/mL manifest extracapsular extension (EPE), and the radial extent of EPE is usually in the range of 2-5 mm [7-9], periprostatic treatment margins accomplished by either monotherapeutic brachytherapy and/or the addition of supplemental external beam radiation therapy (XRT) are necessary to ensure geographic coverage of potential sites of EPE.
Currently, there is interest in minimizing and/or eliminating supplemental XRT in patients with higher risk disease [10]. A reduction/elimination of supplemental XRT will mandate adequate brachytherapy periprostatic treatment margins to address possible EPE and to maximize long term cancer control. At the present time, there is no consensus regarding the extent of periprostatic margins or the dose necessary to sterilize EPE, even among brachytherapy experts [11]. Previously in the Pro-Qura database (Pro-Qura, Seattle, WA, USA), we have documented substantial differences in overall intraprostatic sector dosimetric quality [12]. In the current Pro-Qura evaluation, we evaluate extracapsular dose distributions in patients implanted with permanent prostate brachytherapy. Hopefully, analysis of the Pro-Qura database and other multi-institutional studies will illustrate potential population-based inadequacies and help establish national standards of care. Material and methodsFrom August 1999 to December 2008, 4547 post-implant computed tomography (CT) scans in the Pro-Qura database were available for analysis. The CT scans originated from 129 Pro-Qura participating brachytherapists. Patients implanted at the authors institutions are not part of the Pro-Qura database. The original post-implant prostate dosimetry was reported in terms of a V100 (the percentage of the prostate volume covered by the prescription dose) and D90 (the maximum dose covering 90% of the prostate volume). All implants were pre-planned. Post implant CT was performed at a median of 30 days following brachytherapy.
The Pro-Qura technique for post-implant dosimetric analysis has been described in detail [13]. In this study, using the Pro-Qura defined post-implant prostate volume, a 5 mm, 3-dimensional periprostatic annulus was constructed around the prostate gland (except posteriorly where a 2 mm margin was used), and evaluated in its entirety and in separate sectors to include the anterior, posterior, inferior and right/left apical aspects. Dose to the periprostatic annulus was defined in terms of an annular V100. Figure 1 is an illustration of the periprostatic annulus and the individual segments.
Of the 4547 patients, 3094 (68.0%) were implanted with 125I, 1437 (31.6%) with 103Pd and 16 (0.4%) with 131Cs. For 125I, 84.3% of patients underwent monotherapy (144-145 Gy) and 15.7% a boost (110 Gy) for 103Pd, 67.4% of patients underwent monotherapy (125 Gy) and 32.6% a boost (90-100 Gy). For 131Cs, 62.5% of patients underwent monotherapy (115 Gy) and 37.5% a boost (84 Gy). Because of small patient numbers, 131Cs patients were not included in the analysis of periprostatic treatment margins, but were included in Table 1 for completeness.
Statistical analysis was performed using Predictive Analytics Software (PASW), Statistics Version 17.0 (SPSS Inc., Chicago, IL, USA). The means for continuous variables were compared using independent-samples t-tests, and one-way analysis of variance and chi-square tests were used to compare distributions within categorical variables. Probabilities of deviation from the null hypothesis of no significant differences were marked if statistically significant, p < 5%. ResultsTreatment and summary dosimetric data for the 4547 patients in the study population are summarized in Table 1. There was no statistically significant difference in prostate V100 or D90 when stratified by isotope or monotherapy vs. boost. In addition, there was no difference in prostate size or number of implanted seeds when stratified by isotope. However, for both 125I and 103Pd, prostate glands were statistically larger and more seeds were implanted in the monotherapy vs. the boost cohorts.
Table 2 summarizes the mean margin sector analysis for V100 for the 4547 evaluated patients. For all sectors, the mean V100 was 62.8%. The V100 at the base (51.6%) was less than the apex (73.5%) and the midgland (65.9%). In addition, for the group as a whole, the anterior V100 (45.7%) was less than the lateral (68.8%) and the posterior (75.0%) V100. Tables 3 and 4 describe the mean margin sector volumes, and V100 for 125I and 103Pd. The V100 for 125I was greater than the V100 for 103Pd both overall, and when evaluated by apex/ midgland/base and anterior/lateral/posterior sectors. DiscussionDespite favorable long term biochemical control rates in patients treated with permanent prostate brachytherapy, the definition of a technically adequate implant including periprostatic dose distributions remains somewhat unclear [14,15]. However, data suggests that permanent cancer control is related to intraprostatic radiation dose and periprostatic treatment margins [3-6].
Since extracapsular treatment margins can vary substantially in patients with high quality intraprostatic brachytherapy [5], analysis of annular brachytherapy doses will become mandatory as supplemental XRT is phased out of higher risk brachytherapy protocols. Periprostatic doses are attainable via either a brachytherapy approach that includes generous periprostatic treatment margins and/or the addition of XRT [3,4,11].
Data suggests that periprostatic dose impacts the likelihood of treatment success [5,6,9,16]. Choi et al. demonstrated that the anterior treatment margin in low risk prostate cancer patients was statistically significant in predicting biochemical outcome [5]. In a more sophisticated study, Crook et al. using magnetic resonance imaging (MRI) defined prostate contours reported dosimetric coverage of the prostate gland with 2-, 3-, and 5-mm margins in patients biochemically controlled and in those with biopsy proven recurrence [16]. The mean D90 and V100 were statistically lower in the 2-, 3-, and 5-mm expansions in patients with biopsy proven local failure. Among patients without and with local recurrence, the mean V100 in the 5 mm-group were 77.4% and 72.4% (p = 0.045). Merrick et al. reported an annular V100 of 95.1% in a series of 125I and 103Pd patients [6]. In that study, margin status did not correlate with biochemical control, probably as a result of robust intraprostatic and annular dosimetry. In the current Pro-Qura study, the overall annular V100 was 62.8% for all evaluated patients (65.2% for 125I and 57.6% for 103Pd), which compared to the above mentioned results from high volume brachytherapy centers are probably inadequate for monotherapeutic approaches, especially those with higher risk disease. Previously, it has been reported that 125I resulted in higher annular doses compared to 103Pd [6], but did not result in biochemical control differences.
Although the radiation dose needed to sterilize periprostatic disease is unknown, the dose to control extraprostatic disease is probably significantly less than the threshold intraprostatic doses, because the ratio of extraprostatic to intraprostatic cancer is in the range of 0.4% [17]. Eventually, predictive modeling and improved imaging may enable prostate brachytherapists to tailor treatment margins on a case by case basis [18,19]. Until these technologies become available, a 3-5 mm periprostatic treatment margin appears prudent [6,9,16].
A strength of our analysis is the diverse representation of a large number of community brachytherapy practices with post-implant dosimetry performed with a consistent and highly reproducible technique. However, a limitation of the study is that once stored in the Pro-Qura database, the individual brachytherapist responsible for the CT scan is no longer identifiable and as such a learning curve analysis for margin assessment was not possible. Most importantly, there are limitations to the Pro-Qura post-implant dosimetric technique. Because the Pro-Qura technique uses the pre-implant TRUS determined prostate volume, it is highly probable that the actual annular doses are less than what is reported in this study. In addition, due to the rapid dose fall off at the periphery of the target volume, annular dose distributions are very sensitive to contouring errors. In the current study, this variable has been minimized [8]. Finally, Pro-Qura was established as a preplanning and dosimetry service, and does not have access to outcomes including biochemical control and/or complications. This eliminates our ability to establish a dose response curve for the annular dose necessary to secure long-term biochemical control. ConclusionsThe optimal extracapsular brachytherapy dose and radial extent remains unknown, but will prove increasingly important with reductions and/or elimination of supplemental external beam radiation therapy. The Pro-Qura database demonstrates periprostatic annular doses that are not as robust as those in selected high volume brachytherapy centers, and may be inadequate for optimal biochemical control following monotherapeutic brachytherapy, especially in higher risk patients.Conflict of interestPeter Grimm is an owner of Pro-Qura.References 1. Taira AV, Merrick GS, Butler WM et al. Long-term outcome for clinically localized prostate cancer treated with permanent interstitial brachytherapy. Int J Radiat Oncol Biol Phys 2010; 79: 1336-1342.
2. Zelefsky MJ, Kuban DA, Levy LB et al. Multi-institutional analysis of long-term outcome for stages T1-T2 prostate cancer treated with permanent seed implantation. Int J Radiat Oncol Biol Phys 2007; 67: 327-333.
3. Merrick GS, Butler WM, Wallner KE et al. Dosimetry of an extracapsular anulus following permanent prostate brachytherapy. Am J Clin Oncol 2007; 30: 228-233.
4. Merrick GS, Butler WM, Wallner KE et al. Extracapsular radiation dose distribution after permanent prostate brachytherapy. Am J Clin Oncol 2003; 26: e178-e189.
5. Choi S, Wallner K, Merrick GS et al. Treatment margins predict biochemical outcomes after prostate brachytherapy. Cancer J 2004; 3: 175-180.
6. Bittner N, Merrick GS, Butler WM et al. The correlation between annular treatment margins and biochemical failure in prostate brachytherapy patients with optimized intraprostatic dosimetry. Brachytherapy 2011; 10: 409-415.
7. Partin AW, Mangold LA, Lamm DM et al. Contemporary update of prostate cancer staging nomograms (Partin Tables) for the new millennium. Urology 2001; 58: 843-848.
8. Dawson JE, Wu T, Roy T et al. Dose effects of seed placement deviations from preplanned positions in ultrasound guided prostate implants. Radiother Oncol 1994; 32: 268-270.
9. Davis BJ, Pisansky TM, Wilson TM et al. The radial distance of extraprostatic extension of prostate carcinoma: implications for prostate brachytherapy. Cancer 1999; 85: 2630-2637.
10. Merrick GS, Wallner KE, Butler WM et al. 20 Gy versus 44 Gy of supplemental external beam radiotherapy with palladium-103 for patients with greater risk disease: results of a prospective randomized trial. Int J Radiat Oncol Biol Phys 2012; 82: e449-e455.
11. Merrick GS, Butler WM, Wallner KE et al. Variability of prostate brachytherapy preimplant dosimetry: a multi-institutional analysis. Brachytherapy 2005; 4: 243-251.
12. Merrick GS, Butler WM, Grimm P et al. Multisector prostate dosimetric quality: analysis of a large community database. Brachytherapy 2014, in press. http://dx.doi.org/10.1016/ j.brachy.2013.08.003
13. Merrick GS, Grimm PD, Sylvester J et al. Initial analysis of Pro-Qura: a multi-institutional database of prostate brachytherapy dosimetry. Brachytherapy 2007; 6: 9-15.
14. Aronowitz JN, Rivard MJ. The phylogeny of permanent prostate brachytherapy. J Contemp Brachytherapy 2013; 5: 89-92.
15. Knaup C, Mavroidis P, Stathakis S et al. Evaluation of the effect of the prostate volume change on tumor control probability in LDR brachytherapy. J Contemp Brachytherapy 2011; 3: 125-130.
16. Crook J, Patil N, Ma C et al. Magnetic resonance imaging-defined treatment margins in iodine-125 prostate brachytherapy. Int J Radiat Oncol Biol Phys 2010; 77: 1079-1084.
17. Davis BJ, Haddock MG, Wilson TM et al. Treatment of extraprostatic cancer in clinically organ-confined prostate cancer by permanent interstitial brachytherapy: is extraprostatic seed placement necessary? Tech Urol 2000; 6: 70-77.
18. Ellis R, Zhou E, Fu P et al. Single photon emission computerized tomography with capromab pendetide plus computerized tomography image set co-registration independently predicts biochemical failure. J Urol 2008; 179: 1768-1773.
19. Augustin H, Fritz G, Ehammer T et al. Accuracy of 3-Tesla magnetic resonance imaging for the staging of prostate cancer in comparison to the Partin tables. Acta Radiol 2009; 50: 562-569.
Copyright: © 2013 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License ( http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
|
|